. 2023 Feb 1;13(3):1937–1947. doi: 10.21037/qims-22-618

Table 1. Average visual grading analysis results from four radiologists.

Tube input* ASiR-V DLIR P
30 50 70 Low Medium High
70 kVp, 20 mAs 3.33 [2, 4] 3.50 [2, 4] 3.67 [3, 5] 3.75 [2, 5] 4.08 [3, 5] 4.92 [4, 5] <0.001
70 kVp, 50 mAs 3.42 [3, 4] 3.75 [3, 4] 4.08 [4, 5] 3.75 [3, 5] 4.67 [4, 5] 5.00 [5, 5] <0.001
80 kVp, 20 mAs 3.42 [3, 4] 3.83 [3, 5] 4.33 [4, 5] 3.17 [3, 4] 4.17 [3, 5] 4.75 [4, 5] <0.001
80 kVp, 50 mAs 3.42 [3, 4] 3.83 [3, 5] 4.25 [4, 5] 3.92 [3, 5] 4.58 [3, 5] 4.83 [4, 5] <0.001
100 kVp, 20 mAs 3.67 [3, 4] 3.50 [3, 4] 4.42 [4, 5] 3.92 [3, 5] 4.58 [4, 5] 4.83 [4, 5] <0.001
100 kVp, 50 mAs 3.67 [3, 4] 4.08 [4, 5] 4.17 [4, 5] 4.08 [3, 5] 4.58 [4, 5] 4.83 [4, 5] <0.001
120 kVp, 20 mAs 3.50 [3, 5] 3.83 [3, 4] 4.17 [3, 5] 3.92 [3, 5] 4.58 [4, 5] 4.83 [4, 5] <0.001
120 kVp, 50 mAs 3.75 [3, 4] 4.00 [3, 5] 4.33 [4, 5] 4.33 [4, 5] 4.75 [4, 5] 4.75 [4, 5] <0.001
Total 3.63 [2, 4] 3.93 [2, 5] 4.24 [3, 5] 3.93 [2, 5] 4.54 [3, 5] 4.90 [4, 5] <0.001

Evaluation results are expressed as mean [min, max]. *, CTDIvol values were 0.24, 0.63, 0.38, 0.96, 0.61, 1.53, 0.99 and 2.47 mGy for (70 kVp, 20 mAs), (70 kVp, 50 mAs), (80 kVp, 20 mAs), (80 kVp, 50 mAs), (100 kVp, 20 mAs), (100 kVp, 50 mAs), (120 kVp, 2 mAs), and (120 kVp, 50 mAs), respectively; , highest value. ASiR-V, adaptive statistical iterative reconstruction; CTDIvol, volume computerized tomography dose index; DLIR, deep learning-based image reconstruction.