Skip to main content
. 2022 Dec 13;46(3):392–399. doi: 10.1007/s00270-022-03324-7

Table 2.

Temperature difference of the Bentley/BeGraft after 431 s MPI scan under the influence of different drive field strengths and selection field gradients

Drive field strength (mT) Selection field gradient (T/m, in z-direction) ΔT (K)
3 2.5 0
6 2.5 5.8
9 2.5 16.3
12 2.5 28.0
12 1.25 27.9
12 0 27.9