Skip to main content
. 2023 Feb 20;24(3):e13829. doi: 10.1002/acm2.13829

TABLE 6.

HDR brachytherapy TPS imaging and planning tool validation tests

Test description Tolerance recommended (required) Required
Image transfer and reconstruction Pass/Fail
Patient orientation Pass/Fail
Labeling Pass/Fail
Geometric accuracy Modality‐dependent (see text)
Image registration Modality‐dependent (see text)
Contouring Functional
Source, point, and line delineation 1 mm (2 mm)
External device interfaces (e.g., steppers) Functional