TABLE 7.

HDR brachytherapy TPS source validation and dose calculation tests

Test description Frequency Tolerance recommended (required) Required test
Source model dataI C, A a Exact
Source decay (if possible)II C, SE 1%
Plan normalization/weighting/scalingIII C Functional
Dose calculation gridIV C Functional
Point dose calculation (single source)V C, A a 2% (3%)
Point dose calculation (multisource)V C 3% (5%)
Dose display (absolute and relative)VI C Functional
DVH calculationVII C Functional

Abbreviations: A, annual; C, commissioning; SE, source exchange.

a

Perform either test––see discussion in 8.2.5.