Skip to main content
. 2023 Feb 20;24(3):e13829. doi: 10.1002/acm2.13829

TABLE 8.

HDR brachytherapy TPS commissioning tests

Treatment planning test description Test specifics Tolerance recommended (required) Required test
Optimization validationI Manual dwell time/weight Functional
Dose shaper/graphical optimization Functional
Geometric optimization Functional
Inverse planning Functional
TPS output validationII Printer or pdf function Functional
Data transfer integrity Functional
Applicators and cathetersIII Solid applicator geometry 1 mm (2 mm)
Source position

2 mm (3 mm)

Depends on applicator and modality

Shielding Functional
Independent calculationIV a Dose calculation Functional
Dry run validationV Basic end‐to‐end testing Functional
a