TABLE 8.
Treatment planning test description | Test specifics | Tolerance recommended (required) | Required test |
---|---|---|---|
Optimization validationI | Manual dwell time/weight | Functional | |
Dose shaper/graphical optimization | Functional | ||
Geometric optimization | Functional | ||
Inverse planning | Functional | ||
TPS output validationII | Printer or pdf function | Functional | |
Data transfer integrity | Functional | ✓ | |
Applicators and cathetersIII | Solid applicator geometry | 1 mm (2 mm) | ✓ |
Source position |
2 mm (3 mm) Depends on applicator and modality |
✓ | |
Shielding | Functional | ||
Independent calculationIV a | Dose calculation | Functional | |
Dry run validationV | Basic end‐to‐end testing | Functional | ✓ |
Optional test.