Fig. 2.

Fig 2

Cone-beam computed tomography (CBCT) instrument calibration by K2HPO4 and data analysis by 2 different software products. A, By using a defined concentration series of K2HPO4 (dipotassium hydrogen phosphate) electrolyte solution calibration was attained. The solutions were used in the range of 50 mg/mL to 1000 mg/mL K2HPO4, yielding corresponding Hounsfield unit (HU) values; control reference values were obtained using air and water. The representative CBCT images of these radiographs illustrate the corresponding grey value levels. B, CBCT analysis of the calibration solutions of K2HPO4 was performed by using 2 different software products: blue line = Planmeca Promax 3D (Planmeca Oy) and red line = Simplant 16 (Densply). These 2 analytical tools, based on product specific reconstruction algorithms, delivered identical results, thus providing a basis for data comparison irrespective of the equipment used.