Skip to main content
. 2023 Mar 22;13:4678. doi: 10.1038/s41598-023-31970-5

Table 4.

Percentage of eyes within certain range of prediction error by biometric formula and type of intraocular lenses.

 ± 0.5 (%)  ± 0.75 (%)  ± 1 (%)
AL ≤ 22 mm
 SRK/T Monofocal 84.6 100 100
Multifocal 50.0 100 100
 Haigis Monofocal 75.0 100 100
Multifocal 100 100 100
 Hoffer Q Monofocal 76.9 100 100
Multifocal 100 100 100
 Holladay I Monofocal 84.6 100 100
Multifocal 100 100 100
 Kane Monofocal 91.7 100 100
Multifocal 100 100 100
 Barrett Monofocal 84.6 100 100
Multifocal 100 100 100
22 < AL < 25 mm
 SRK/T Monofocal 87.8 96.3 100
Multifocal 90.0 97.5 100
EDOF 81.1 94.6 100
 Haigis Monofocal 77.9 95.8 100
Multifocal 92.1 97.4 100
EDOF 45.4 78.7 100
 Hoffer Q Monofocal 85.0 97.9 100
Multifocal 95.0 97.5 100
EDOF 86.5 97.3 100
 Holladay I Monofocal 86.5 99.3 100
Multifocal 95.1 97.5 100
EDOF 89.2 97.3 100
 Kane Monofocal 81.6 97.2 100
Multifocal 92.1 100 100
EDOF 78.9 92.1 100
 Barrett Monofocal 82.8 96.6 100
Multifocal 85.0 100 100
EDOF 78.9 92.1 100
AL ≥ 25 mm
 SRK/T Monofocal 89.5 100 100
Multifocal 100 100 100
 Haigis Monofocal 94.4 100 100
Multifocal 85.7 100 100
 Hoffer Q Monofocal 94.4 100 100
Multifocal 100 100 100
 Holladay I Monofocal 84.2 100 100
Multifocal 100 100 100
 Kane Monofocal 94.4 100 100
Multifocal 85.7 100 100
 Barrett Monofocal 89.5 100 100
Multifocal 87.5 100 100