Comparison of the accuracy and optimal cutoff values in judging the anesthesia-induced unconsciousness, LOC, and ROC between Ai and BIS.

States of consciousness Index AUC SE 95%CI Cutoff value Yd
Anesthesia-induced Ai 0.941 0.00401 0.933–0.948 71.5 0.789
Unconsciousness BIS 0.943 0.00412 0.935–0.950 72.5 0.814
LOC Ai 0.953 0.00437 0.944–0.960 79.5 0.882
BIS 0.965 0.00424 0.957–0.971 79.5 0.855
ROC Ai 0.936 0.00456 0.927–0.944 70.5 0.799
BIS 0.934 0.00479 0.925–0.942 71.5 0.778