. 2023 Feb 28;61(3):e01525-22. doi: 10.1128/jcm.01525-22

FIG 3.

FIG 3

Accuracy of the ASTar MIC compared to the reference MIC, with the number of doubling dilutions from the reference MIC. In most cases, the reference MIC range covers or exceeds the full ASTar range. For a subset (6.6%) where reference MICs do not cover the full range of the MICs of the ASTar BC G kit, the comparisons of ASTar MIC ranges have been shortened to be comparable with the reference MIC ranges.