. 2023 Mar 27;7:15. doi: 10.1186/s41747-023-00329-w

Table 4.

SNR and CNR calculations

Kernel Position Dose protocol PCD-CT (mean ± SEM) EID-CT (mean ± SEM) p value
SNR Br76/Ur77 OH 2 mGy 0.6 ± 0.11 0.3 ± 0.05 < 0.001
6 mGy 0.7 ± 0.11 0.3 ± 0.04 < 0.001
12 mGy 0.7 ± 0.11 0.6 ± 0.15 0.001
ABD 2 mGy 0.5 ± 0.16 0.3 ± 0.05 < 0.001
6 mGy 0.7 ± 0.16 0.3 ± 0.05 < 0.001
12 mGy 0.7 ± 0.17 0.4 ± 0.07 < 0.001
Br89/Ur89 OH 2 mGy 0.3 ± 0.07 0.3 ± 0.01 0.031
6 mGy 0.4 ± 0.09 0.3 ± 0.02 < 0.001
12 mGy 0.4 ± 0.12 0.2 ± 0.03 < 0.001
ABD 2 mGy 0.3 ± 0.58 0.4 ± 0.03 < 0.001
6 mGy 0.3 ± 0.08 0.3 ± 0.01 0.235
12 mGy 0.3 ± 0.11 0.3 ± 0.02 0.774
CNR Br76/Ur77 OH 2 mGy 0.9 ± 0.16 0.4 ± 0.04 < 0.001
6 mGy 1.0 ± 0.14 0.6 ± 0.05 < 0.001
12 mGy 1.0 ± 0.15 1.0 ± 0.12 0.013
ABD 2 mGy 0.5 ± 0.32 0.4 ± 0.10 < 0.001
6 mGy 0.9 ± 0.20 0.4 ± 0.04 < 0.001
12 mGy 0.9 ± 0.21 0.5 ± 0.06 < 0.001
Br89/Ur89 OH 2 mGy 0.4 ± 0.11 0.4 ± 0.03 0.031
6 mGy 0.6 ± 0.13 0.3 ± 0.02 < 0.001
12 mGy 0.7 ± 0.18 0.3 ± 0.04 < 0.001
ABD 2 mGy 0.4 ± 0.71 0.4 ± 0.04 < 0.001
6 mGy 0.4 ± 0.11 0.1 ± 0.03 < 0.001
12 mGy 0.5 ± 0.13 0.4 ± 0.02 0.011

Calculations of SNR and CNR after a ROI-based measurement of signal attenuation for different dose protocols, arm scanning positions and reconstruction kernels. The results are given as mean values with the standard error of mean (± SEM). The p value of the Mann-Whitney U test analysis is reported. ABD Abdominal scanning position, CNR Contrast-to-noise ratio, EID-CT Energy-integrating detector computed tomography, OH Overhead scanning position, PCD-CT Photon-counting detector computed tomography, ROI Region of interest, SNR Signal-to-noise ratio