. 2022 Dec 22;4(3):207–214. doi: 10.1016/j.hroo.2022.12.010

Table 3.

Odds of CIED infection from prespecified risk factors

Predictor Adjusted odds ratio (95% CI) P value
Type of current device .041
PPM 1.0 (reference)
AICD 0.4 (0.1–1.0)
CRT 2.3 (0.5–11.0)
Organism of bacteremia <.001
CoNS 19.0 (4.5–80.7)
Enterococci 13.8 (3.7–52.0)
VGS 14.7 (3.4–63.9)
Other non-SA GPC 1.0 (reference)
Duration of bacteremia (per 1 d of bacteremia) 2.1 (1.3–3.2) .001
Type of bacteremia <.001
Community acquired 1.0 (reference)
Healthcare associated 0.1 (0.04–0.4)
Nosocomial 0.2 (0.1–0.7)

Results obtained from a multivariable logistic regression model for CIED infection based on all the variables listed in the table.

AICD = automatic implantable cardioverter-defibrillator; CI = confidence interval; CIED = cardiovascular implantable electronic device; CoNS = coagulase-negative Staphylococcus; CRT = cardiac resynchronization therapy; ESRD = end-stage renal disease; GPC = gram-positive cocci; SA = Staphylococcus aureus; VGS = viridans group streptococci.