Skip to main content
. 2023 Mar 28;23:179. doi: 10.1186/s12903-023-02873-8

Table 3.

Implant deviations using Yizhimei dynamic navigation system

Deviation Max Min Mean SD 95%CI
Lower Upper
Platform (mm) 1.12 0.49 0.73 0.20 0.58 0.87
Apex (mm) 1.35 0.40 0.86 0.33 0.62 1.09
Angulation° 3.74 1.66 2.32 0.71 1.81 2.82