Robustness assessment of the developed TLC–densitometric methods for determination of IMD and DLM

Robustness parameter SD of peak area % RSD
IMD DLM IMD DLM
(1) Volume of mobile phase
 98 ml 144.07 100 2.87 2.78
 100 ml 85.26 31.18 1.68 0.83
 102 ml 122.46 40.40 2.48 0.97
(2) Duration of saturation
 25 min 296.45 26.93 1.46 1.43
 30 min 227.33 11.89 1.11 0.61
 35 min 297.58 37.55 1.60 1.97