Skip to main content
. Author manuscript; available in PMC: 2023 Apr 5.
Published in final edited form as: Curr Protoc. 2022 Dec;2(12):e614. doi: 10.1002/cpz1.614

Table 7.

Troubleshooting CRISPR-suppressor scanning experiments

Problem Possible Cause Solution
Insufficient number of colonies is observed following library cloning transformation.
  • Gibson Assembly reaction was low efficiency.

  • Transformed DNA is low quality.

  • Electroporation was low efficiency.

  • Insufficient cells were electroporated.

  • Further purify sgRNA inserts and/or digested lentiCRISPRv2 vector using a PCR purification kit to remove any contaminates left over from gel purification.

  • Ensure no residual ethanol from the wash is present when resuspending sample at the end of the isopropanol precipitation.

  • Adjust electroporation conditions used to ensure that cuvette does not arc, and minimize lag time between electroporation and resuspension of the cells in recovery medium.

  • Scale up the number of cells electroporated by performing additional electroporations.

Sequenced sgRNA library shows poor statistics (e.g., many undetected sgRNAs, high skew ratio).
  • Errors or biases were introduced during PCR amplification steps.

  • Optimize cycle numbers for rounds 1 and 2 of PCR amplification for sgRNA library cloning, ensuring that less than 35–40 cycles are used in total across both amplifications.

  • Optimize cycle numbers used in library amplification for NGS analysis, testing cycle numbers less than the recommended 22 cycles.

  • Ensure that a high-fidelity DNA polymerase is used, such as Q5 hot-start high fidelity DNA polymerase.

Low lentiviral titer is observed.
  • Poor quality HEK293T cells were used for lentivirus production.

  • Poor quality DNA was used for transfection.

  • Non-optimal transfection conditions were used.

  • Virus was handled improperly.

  • Non-optimal transduction conditions were used.

  • Cell line of interest is difficult to transduce.

  • Ensure that HEK293T cells are low passage (passage <10) and do not exceed 70–80% confluency at any time prior to seeding for transfection.

  • Verify the quality of pMD2.G and psPAX2 plasmids by sequencing or diagnostic digest and repurify if necessary.

  • For transfection, ensure that FuGENE HD is pipetted directly into the liquid in the tube and not against the tube walls, and limit incubation to 10–15 min.

  • Optimize the volume of FuGENE HD used for transfection relative to the total amount of DNA.

  • Ensure that lentivirus is snap-frozen rapidly on dry ice and is not subjected to additional freeze-thaw cycles. Thaw lentivirus slowly on ice.

  • Concentrate lentivirus upon harvesting and prior to freezing, especially if the cell line of interest is difficult to transduce.

  • Optimize transduction conditions by adjusting the number of cells seeded per well, the volume of lentivirus, and the polybrene concentration. Optimal conditions will vary for each cell line.

  • Scale up the number of cells transduced to achieve coverage.

Number of cells transduced during the screen is unexpectedly high or low.
  • Cells were handled inconsistently between titering transduction and screen transduction.

  • Lentivirus aliquots became degraded.

  • Ensure that cells are handled the same way, and that transductions are performed using identical conditions, between titering and screening experiments.

  • Calculate the number and proportion of cells successfully transduced. If these metrics are within acceptable MOI and coverage recommendations, proceed with the screen.

  • Ensure that lentivirus used for titering is treated the same way as that used for the final screen (i.e., subjected to a freeze-thaw cycle, thawed on ice, etc.).

  • Avoid subjecting lentivirus aliquots to multiple freeze-thaw cycles.

Low viability of cells is observed following drug selection during CRISPR-suppressor scanning experiment.
  • Too high of a drug concentration was used.

  • Reduce the concentration of drug used (i.e., use a less stringent selection).

Analysis of CRISPR-suppressor scanning results shows high replicate variability or unexpected behavior of controls.
  • Adequate coverage of the sgRNA library was not achieved. For example, cell populations were bottlenecked during the screen or an insufficient amount of DNA was harvested and amplified.

  • Optimize transduction conditions and carefully monitor cell counts during the screen to ensure that adequate coverage (ideally 1,000× coverage of the library) is maintained at all times.

  • Scale up the number of cells subjected to drug treatment or reduce the concentration of drug used.

  • Harvest an excess number of cells to ensure sufficient DNA is extracted.