Table 2.

Electrochemical response characteristics and validation parameters of the proposed Midazolam sensors

Parameter PTFANI-free sensor PTFANI-sensor
Slope (mV/decade)a 53.5 54.5
Intercept (mV)a 301.0 332.4
Correlation coefficient (r) 0.9992 0.9998
Response time (s) 25 10
Working pH range 2.5–4 2.5–4
Concentration range (M) 1 × 10–6–1 × 10–2 1 × 10–6–1 × 10–2
Stability (days) 20 30
Accuracy (mean ± SD) 100.13 ± 1.57 99.89 ± 1.13
Precision (% RSD)
 Intra-day precisiona 0.864 1.311
 Interday precisionb 1.296 1.967
 LOD (M)c 9.0 × 10–7 9.0 × 10–7

aThree different concentrations of three replicate each (n = 9) repeated three times within the same day

bThree different concentrations of three replicate each (n = 9) repeated on three successive days

cLimit of detection (measured by intersection of the extrapolated arms of non-responsive and the Nernstian segments of the calibration plot