TABLE 2.

Meta‐analysis of risk factors for cardiac implantable electronic devices malfunctions associated with radiation therapy

Studies (n) Patients (n) Malfunctions (n) Risk ratio (95% CI) p I 2% (95% CI)

Egger's test

p

Neutron‐producing vs. non‐neutron‐producing energy 13 1350 65 9.98 (5.09–19.60) <.0001 4 (0–61.7) .472
PM/ CRT‐P vs. ICD/ CRT‐D 25 3018 123 2.07 (1.40–3.06) .0003 1 (0–49.4) .074
Dmax ≤2Gy vs. Dmax >2Gy 8 718 17 0.93 (0.31–2.76) .8983 0 (0–84.7) .282

Abbreviations: CRT‐D, cardiac resynchronization therapy and defibrillator; CRT‐P, cardiac resynchronization therapy and pacing; Dmax, maximal radiation dose at device; Gy, grey; ICD, implantable cardioverter defibrillator; PM, pacemakers.