. 2022 Sep 29;36(1):e4825. doi: 10.1002/nbm.4825

TABLE 1.

Image acquisition parameters

Figure Sample / Scanner Sequence FOV (mm3) No of pixels Bandwidth (kHz) TR (ms) TE (ms) Other params Avgs Scan time (min)
 6A) Phantom / i3M 3D‐SE
130×130×130
60×60×30
30 500 20 1 15
 6B) Phantom / i3M 3D‐RARE
130×130×130
60×60×30
30 500 20 Center‐out ETL = 10 ESP = 20 ms 1 1.5
 6C) Phantom / i3M 3D‐STIR
130×130×130
60×60×30
30 500 20 TI = 20 ms 4 60
 6D) Phantom / i3M 3D‐GRE
130×130×130
60×60×30
60 10 2 Flip angle =  15 1 9
 6E) & F) Phantom / i3M 3D‐SE (radial)
130×130×130
60×60×30
30 100 20 Radial in y,z Equispaced in x Spokes per slice = 95 1 4.8
 7 Phantom / LUMC 3D‐SE
150×150×150
100×100×20
20 1000 30 1 33.3
 8 Phantom / MGH 2D‐RARE
15×15
128×128
30 5000 53 Axial ETL = 32 ESP = 9.5 ms 4 1
 9A) & E) Elbow / i3M 3D‐RARE
120×120×120
120×120×25
30 400 20 Center‐out ETL = 10 ESP = 20 ms 6 12
 9B) & F) Forearm / i3M 3D‐RARE
120×120×100
120×120×25
30 400 20 Center‐out ETL = 10 ESP = 20 ms 6 12
 9C) & G) Wrist / i3M 3D‐RARE
200×160×80
120×120×20
30 400 20 Center‐out ETL = 10 ESP = 20 ms 7 11.2
 9D) & H) Hand / i3M 3D‐RARE
180×180×50
120×120×10
30 400 20 Center‐out ETL = 10 ESP = 20 ms 13 10.4
 10A) & D) Ankle / i3M 3D‐RARE
160×180×120
80×90×10
20 100 20 Center‐out ETL = 3 ESP = 20 ms 36 18
 10B) & E) Ankle / i3M 3D‐RARE
160×180×120
80×90×10
20 1000 40 Out‐out ETL = 3 ESP = 40 ms 2 9
 10C) & F) Ankle / i3M 3D‐RARE
160×180×120
80×90×10
20 1000 20 Center‐out ETL = 5 ESP = 20 ms 3 14

Note: TE in RARE sequences corresponds to the effective echo time.