Abstract
The use of volumetric arc therapy and inverse planning has been in routine use in radiotherapy for two decades. However, use in total body irradiation (TBI) has been more recent and few guidelines exist as to how to plan or verify. This has led to heterogeneous approaches. The goal of this review is to provide an overview of current advanced planning and dosimetry verification protocols used in optimised conformal TBI as a basis for investigating the need for greater standardisation in TBI.
Introduction
Total body irradiation (TBI) is a well-established external beam radiotherapy treatment employed as part of the conditioning regimen prior to bone marrow, haematopoietic or peripheral blood progenitor stem cell transplantation with the aim to supress the patient’s immune system and prevent rejection of the donor’s stem cells. 1,2 High dose, myeloablative TBI treatments should also eradicate malignant tumour cells, particularly in sanctuary sites which chemotherapy may find more difficult to penetrate, such as the brain or the testes. Patients undergoing myeloablative TBI are at risk of significant toxicities, in particular to the lungs with the risk of developing interstitial pneumonitis.. 2,3 Patients who are unable to tolerate myeloablative TBI undergo lower dose single fractionation regimes (reduced conditioning or non-myeloablative TBI). 2 .
Radiation treatment consists of the delivery of high megavoltage (MV) energy photons to the entire body. Conventional TBI is usually delivered using a linear accelerator at an extended source-to-surface distance (SSD) so that the treatment field may encompass the entire patient, whether in a standing, lying or semi-seated position. At least two opposing fields are delivered, e.g. using a right- and left lateral for a semi-seated position. To achieve dose homogeneity within ± 10% (as specified in the American Society for Radiation Oncology (ASTRO) practice guideline 4 ) throughout the patient, compensators are used. These consist of either lead or Perspex shielding, water equivalent bolus and/or multileaf collimators within the treatment field. In vivo dosimetry is recommended for TBI to verify the delivery of the prescription dose and doses to organs at risk (OARs). 5 Methods to optimise 3D dose distributions and minimise doses to OARs could confer benefits to these patients and in particular paediatric cases. 2
Optimised conformal TBI techniques are now being implemented in centres worldwide and there has been an upsurge in publications, particularly for the use of isocentric-based techniques such as volumetric arc therapy (VMAT) 6–9 and helical tomotherapy (HT). 10–14 These techniques bring many challenges due to the long treatment volume >140 cm, including planning protocol development, imaging verification, dose rate considerations and dosimetric verification of the treatment plan, which is critical to ensure accurate delivery. Elongated treatment fields require multiple arcs and this leads to overlapping junction regions where the dose needs to be carefully controlled. Modern treatment planning systems (TPSs) offer ways to limit the junction dose, either through the planning method or during optimisation of the plan. Junction regions need to be verified dosimetrically prior to treatment, however, this can be done alongside the standard patient-specific quality assurance (PSQA) performed for VMAT plans.
In paediatric TBI, decreasing doses to OARs is of paramount importance to reduce acute toxicity and late effects. The Children’s Oncology Group (COG) in the United States conducted a recent survey 15 in which 79% of respondents supported investigation of new techniques to lower lung doses and 83% supported the use of VMAT or HT. The European SIOPE survey in 2021 16 concluded that modern TBI techniques with integrated shielding would allow for the reduction of OAR doses. Recent guidelines for paediatric TBI 2 have concluded that myeloablative TBI performance for children is currently heterogeneous and centre-specific.
Introducing advanced techniques
Previous reviews of TBI practice support evaluation of optimised conformal techniques. 17–19 Ideally, the introduction of advanced techniques should form part of a clinical trial, as stated in the International Lymphoma Radiation Oncology Group (ILROG) 2018 guidelines. 1 Wong et al 20 published a detailed review of total marrow irradiation (TMI) and total nodal irradiation (TNI) covering clinical trial results, review of technologies used for treatment (e.g. VMAT) and considered the barriers for implementation of VMAT and HT for TMI/TNI. They suggested that, to increase the uptake of TMI/TNI techniques, the first step should be to introduce intensity modulated radiotherapy (IMRT) for TBI and consider using artificial intelligence (AI) for contouring. Recent papers on VMAT TBI have also considered the use of planning automation 21,22 and flattening filter free delivery 23 to aid implementation of optimised conformal techniques.
Recent national TBI surveys across Europe, Canada, Japan, New Zealand, and the United States 24–28 demonstrated that the majority of centres are still utilising an extended SSD technique with some use of field-in-field (FIF) techniques to improve dose homogeneity. These surveys highlighted multiple areas of heterogeneous practice, in particular the wide range of dose rates in use (2.25–40 cGy/min) and lung tolerance doses (8–12 Gy). Prescription points were only specified in 2/7 surveys and location varied from the umbilicus to an average of midline points. The use of dose–volume histogram (DVH) constraints in CT planning were only discussed in half the surveys. This provides an incomplete picture of the planning protocols in use.
If optimised conformal techniques are implemented as part of a clinical trial, accurate dosimetry is vital to report trial outcomes. 15 External dosimetry audits are an important part of safety culture in radiotherapy. 29–34 They are recommended as best practice for new equipment and techniques, and ensure dosimetric accuracy is maintained and improved. They provide assurance of accuracy for clinical trial participation and support implementation for optimised conformal techniques. The requirement for audit is specified in various recommendations including the NHS England External Beam Radiotherapy Specifications 2019 (NHSE EBRT 2019). 35 TBI dosimetry audits are uncommon and, at the time of writing, there are no full papers published describing dosimetry audits on TBI in the literature. The Imaging and Radiation Oncology Core (IROC) Houston QA centre (Houston, Texas) have performed a paediatric TBI audit using a custom-made phantom simulating a 10-year-old. The audit was conducted remotely at 20 institutions to propose credentialing criteria. This confirmed the wide variety of TBI practices and lung doses in use. 36
Recently published guidelines covering planning and delivery of TBI from ASTRO, 3 ILROG, 1 the Australian College of Physical Scientists and| Engineers in Medicine (ACPSEM) 37 and the Netherlands Commission on Radiation Dosimetry (NCS) 38 , however, do not include complete guidance on planning protocols, volume definition, OAR dose-volume constraints and PSQA.
The development of TBI practice is rapidly evolving to incorporate optimised conformal planning techniques. Implementation is taking place independently within individual institutions, which has led to heterogeneous approaches in treatment planning and delivery. The aim of this review is, therefore, to evaluate the published planning and dosimetry protocols used for optimised conformal TBI to support standardisation of TBI practice.
Current status of planning techniques
A literature search was conducted on the 23 April 2022 using PUBMED and the criteria included were total body irradiation AND planning within the last 10 years (2013–2022) due to the upsurge in publications of optimised TBI. This returned 364 papers, which were distilled down to 88 papers that referred to TBI planning. Papers had to contain details of the TBI planning protocol, including planning target volume (PTV) and OAR dose-volume constraints/objectives, to be included. Papers describing clinical results and protocols for an optimised conformal TBI technique as part of a planning study were also included. Conference abstracts, simulation planning studies with no clinical data or papers covering TMI only were excluded. 28 papers remained, which were also examined for details of PSQA and/or in vivo dosimetry measurements performed as part of clinical practice.
These papers covered a range of techniques, as shown in Table 1. Four papers compared two TBI techniques giving a total of 32 protocols. HT (including TomoDirect) 10–14,39–45 is the most widely used technique followed closely by VMAT 6–9,46,47 or VMAT hybrid techniques, where the lower limb fields are delivered using a FIF anteroposterior/posteroanterior (AP/PA) technique. 48–51 Extended SSD treatments such as FIF, IMRT 52–55 and arc or modulated arc therapy 56,57 also feature. Arc techniques require the patient to be placed on a specially designed couch near to the treatment floor with a supine then prone delivery of multiple arcs. Only one institution delivered VMAT using this setup. 48 Comparing the techniques against year of publication it is evident that papers on VMAT implementation are on the rise (Figure 1). 13/32 of the protocols included paediatric patients, 3 were paediatric only. Modulated approaches were used in all cases for both paediatric and adult patients.
Table 1.
Frequency of techniques used
| Technique | Number of protocols |
|---|---|
| Extended SSD FIF | 3 |
| Extended SSD IMRT | 3 |
| Arcs and/or static at floor | 1 |
| Modulated arc | 1 |
| VMAT near floor | 1 |
| VMAT and FIF | 2 |
| VMAT | 9 |
| HT | 12 |
FIF, field-in-field; HT, helical tomotherapy; IMRT, intensity modulated radiotherapy; SSD, source-to-surface distance; VMAT, volumetric arc therapy.
Figure 1.
Frequency of papers published by technique.
The findings have been summarised according to the ESTRO RATING 58 methodology for reporting treatment planning results and protocols. Areas of common practice, used in more than half the protocols, were limited but included the use of head-first supine (HFS) and feet-first supine (FFS) CT scans due to limited scan length, the use of 6 MV and defining lung as a critical OAR.
CT simulation
For the majority of VMAT and HT techniques, patients had full body CT scans with typical slice thickness of 5 mm, two scans (HFS and FFS), then either fused or planned as two separate plans (16/32). Alternatively, patients were scanned to mid-thigh supine and the dataset extended from the most inferior slice to represent the patient’s full height (2/32) or scanned with bent legs or to mid-thigh (7/32). Arms were placed along the patients’ sides in the majority of protocols (24/32). One paper referred to full body scanning but it is unclear how these were conducted, and four papers did not state the scanning method. When immobilisation techniques are mentioned, vacbags are the main equipment employed (7/32) to ensure setup consistency during HFS and FFS rotation, as well as thermoplastic masks around head and shoulders.
PTV definition
PTV definitions differ. Over two-thirds of the protocols (21/32) based the PTV definition on the whole-body contour with a 3–5 mm margin cropped away from the skin to account for modelling inaccuracies and to avoid overdosing at the skin surface, whereas 5/32 reported the whole body for the volume definition. The inclusion or exclusion of OARs from the final PTV was variable with only half the protocols (16/32) excluding them. The OARs excluded also varied making it difficult to compare PTV dose–volume statistics directly. The most widely used PTV dose-volume constraint was D95 >95% followed by inspection of dose coverage between 80 and 95% of the prescribed dose. These parameters are, of course, directly influenced by the PTV definition. PTV definition and plan prescription for external beam radiotherapy also need to follow standardised protocols to allow comparison of plan dosimetry (ICRU 83 59 ).
PTV coverage/dose–volume objectives
Planning PTV objectives vary widely across protocols. Approximately, half the protocols set hard constraints and either used dose–volume parameters or a visual inspection of dose coverage. The most frequently used dose–volume objectives are given in Table 2, with D95 >95% of the PTV being the most common.
Table 2.
PTV coverage and dose-volume objectives
| PTV dose objectives | Protocols |
|---|---|
| D95 > 95% | 8 |
| Coverage > 95% | 2 |
| Coverage > 90% | 6 |
| Coverage > 85% | 1 |
| Coverage > 80% | 2 |
| D98 90% | 3 |
PTV, planning target volume.
Organs at risk
Lung
The lungs are critical OARs due to the risk of developing interstitial pneumonitis (IPS) at around 8–10 Gy, 1 except from low dose conditioning indications. The main dose constraint reported was mean lung dose (MLD), ranging from 7 to 12.5 Gy, again confirming a lack of consensus. Seven protocols used a MLD constraint <10 Gy, nine a MLD of 10–12.5 Gy, and one reported a range of 8–10 Gy, depending on prescription dose. Other lung dose constraints used included V10Gy <10%, 43 V10Gy <40%, 40 V8Gy >90% with Dmedian <9 Gy 42 and V6Gy >90%, V8Gy <40%. 55 MLD >8 Gy has also been associated with a decrease in overall survival 60 and recent paediatric TBI guidelines recommend limiting MLD to <8 Gy. 2
Kidneys
Kidneys were the second most prevalent OAR with dose-volume constraints stated in 10 of the papers, with a mean dose limit to combined kidneys between 7.5 and 13 Gy. For paedatric TBI a mean dose <10 Gy is recommended. 2
Other OARs
TBI treatments are also known to cause toxicity to organs other than the lungs and kidneys, including the brain, eyes, (with example constraints Dmean < 12 Gy 54 and Dmax <4–6 Gy, respectively 6,8 ) and liver. For paediatric TBI, reducing the lens dose to <12 Gy is recommended. 2
Long-term risks
In the paediatric setting, the increased ability to conform treatments could have benefits for reducing other important long-term effects such as secondary malignancies, infertility, cardiovascular disease, learning deficits and growth failure. 18,60–62 In recent years, the radiosensitivity of the heart has become an increasing area of research, to predict risks of cardiovascular disease. 63 The risk of premature death following haemopoietic stem cell transplant (HSCT) carries a 2.3-fold increased risk due to cardiovascular complications compared to healthy individuals (including TBI). 64 Cardiac doses can be reduced even when using conventional shielding, as highlighted in a recent paper by Glenn et al, 65 although 3D optimisation could take this further. Paix et al 18 state that HT for TMI vs conventional TBI provides a mean dose reduction of between 46 and 47%. Even with these dose reductions, optimised conformal TBI included in HSCT regimes needs to be compared to HSCT regimes using conventional TBI delivery techniques to assess their true effectiveness in terms of dose-toxicity and outcomes and, ideally, as part of a clinical trial. 20
Treatment machines, TPS and energy
The treatment machines used for delivery were stated in 23/32 protocols. HT was the most frequently used machine (10/32) with the Siemens linear accelerator the least frequent. Consequently, the most commonly used TPS was HT (14/32). For VMAT and IMRT-based techniques, Eclipse and Pinnacle were the most widely used TPSs (9/32 and 7/32 respectively). For VMAT techniques, the number of isocentres ranged from 3 to 15 and the number of arcs planned at each isocentre varied between 2 and 8 arcs. Energies were stated in approximately half of the protocols, with 6 MV the most frequently used (range 6–18 MV).
Where dose rates were stated, these were reported in MU/min or cGy/min in relation to the beam or patient, respectively. Reported dose rates ranged from 0.1 to 19 cGy/min for extended SSD techniques or 20 MU/min to 600 MU/min for VMAT/HT techniques. Absorbed dose rates were reported at a range of locations, when specified, including mid-plane and in lung. In some cases, the mean instantaneous dose rate was reported.
Extended SSD TBI techniques, tended to report a dose rate of 300 MU/min with one (step and shoot IMRT) reporting 100 MU/min and another (FIF+VMAT) where the legs were irradiated with 600 MU/min fields.
7/32 protocols using modulated VMAT techniques only reported the average dose rate, while 6/32 protocols made the distinction of lower dose rates being used on arcs irradiating the lungs (i.e. 40 or 60 MU/min for adults and 20 MU/min for children). The rest used 300 MU/min, on average.
Dose rate was not usually reported for HT techniques. Instead, average beam-on time or modulation factor and/or pitch were discussed: The longer the treatment time (within reason for patient’s comfort) the lower the dose rate, and the more homogenous the plans.
For VMAT/HT, 3D dose rate mapping from the TPS could be advantageous as this would enable future analysis of planning dosimetry data vs toxicity data.
Dose rate is often cited as a risk factor for IPS 3,19 and, hence, moving to isocentric techniques, where dose rates are much higher, could increase IPS incidence. Vogel et al 3 published a critical review of pulmonary toxicity after TBI and recommended that a dose rate <15 cGy/min and a MLD <600 cGy was supported by the literature. This conflicts with other papers that recommend a MLD<8 Gy. However, this detailed review of over 100 papers not only found inconsistencies in definitions of IPS but also in the reporting of TBI parameters limiting detailed research into their correlation. Nevertheless, with incomplete reporting of TBI parameters and the lack of clarity on dose rate definition further dosimetric evaluation will not be possible until TBI reporting has reached a sufficient level of standardisation. 2 Only then will it be possible to undertake a comprehensive assessment comparing lung dose–volume statistics, dose rate and IPS incidence and severity.
Fractionation
A wide range of fractionations were used as shown in Figure 2. 11 different fractionations were used across 24 centres. The most frequently used fractionation was 12 Gy in 6 fractions, reported in 5 publications. The most commonly used dose per fraction (#) was 2 Gy, reported by over half of the papers and ranging from 0.83 to 3 Gy. Fractionation, and therefore prescribed doses, varied widely resulting with total doses up to 12–14.4 Gy for high-dose regimes.
Figure 2.
Frequency of fractionations for myeloablative or non-myeloblative and reduced intensity regimes.
Junction dosimetry
Optimised conformal TBI planning comes with practical challenges, particularly in delivering a uniform dose to a PTV > 140 cm whilst minimising junction doses. VMAT and HT full body TBI isocentric techniques require optimisation of the junctions between arcs or superior and inferior plans. There is variation in how institutions handle this region. Of the 21 relevant papers, 17 presented details of how dosimetry was approached (Table 3). For VMAT planning, the majority of centres used either an additional beam or overlap between arcs which the TPS would use to account for the dose distribution during optimisation. For HT, the approach was distinctly different, with PTV subvolumes being created in the overlap region, usually the mid-thigh, with different optimisation objectives to create a dose gradient in opposite directions for the superior and inferior plans. This meant that, on summation, a uniform junction dose could be achieved. Figure 3 shows an example of the subvolumes. 11 Only one institution used PTV subvolumes for their VMAT technique.
Table 3.
Junction optimisation for VMAT and HT plans
| Junction optimisation method | Technique | Number of Protocols |
|---|---|---|
| PTV subvolumes | HT (3), VMAT (1) | 4 |
| PTV amendment to allow for junction volumes | HT | 1 |
| Manual adjustment | HT | 3 |
| TPS optimised with arc overlap | VMAT | 5 |
| TPS FIF MLC optimisation | VMAT | 1 |
| TPS optimised using isodose structure | VMAT | 1 |
| TPS optimised additional beam | VMAT | 2 |
FIF, field-in-field; HT, helical tomotherapy; MLC, multileaf collimator; PTV, planning target volume; TPS, treatment planning system; VMAT, volumetric arc therapy.
Figure 3.
PTV subvolumes (reproduced with permission 11 )
Recent guidelines for paediatric TBI state that field junctions should be evaluated as part of departmental QA procedures and planned dose homogeneity within the target should be within 90–110% of the prescription dose. Few centres stated their acceptable junction dose range (3/17; 14%). There was also minimal reporting on measurement and verification of the junction region for PSQA 44,46 (2/17).
Patient-specific quality assurance (PSQA) and in vivo dosimetry
The final aspect to consider for implementation of optimised conformal techniques includes verification of the treatment plan delivery. A large number of methods and equipment were reported to verify TBI treatments, however, from the papers analysed, only half commented on how PSQA was performed. The majority (11/19) of centres took an absolute dose measurement to verify the dose from the TPS, with other methods including array or in vivo measurements. The remainder did not report any PSQA or in vivo measurements. Several phantoms were used across the centres including solid water, TomoTherapy “Cheese phantom” and Alderson in combination with an ionisation chamber (NACP parallel plate chamber, A1SL), (Figure 4). Other methods used for PSQA included a 2D or 3D array such as ArcCheck, Delta4 and Multicube, where gamma analysis was performed to evaluate the dose distribution. Of nine centres using phantom based PSQA, six did not state the pass rate criteria for gamma analysis. However, the remaining three centres used a 3%/3 mm gamma pass rate criterium with two of the centres passing the plan over 90% and the last centre using a stricter criterium of over 95%. It was not stated how the gamma pass rate was decided, or any other criteria used. 66 According to the recommendations in AAPM TG218, 67 the suitable gamma criteria to assess a standard VMAT plan is 3%/3 mm, but the guidance does not specify any recommendations for TBI treatments.
Figure 4.
Phantoms used for PSQA
Finally, some centres aimed to confirm delivered dose in vivo using TLDs, OSLD, MOSFET and portal system dosimetry. The majority of the in vivo dosimeters were positioned in the thorax and abdomen (but varied in position between centres) with a maximum tolerance of 10% compared to the calculated dose. Within the study comparison, half of the centres (13/33) did not report their method for verifying the dose and several papers stated more than one method of PSQA and/or in vivo dosimetry. In vivo dosimetry using TLD, OSLD or MOSFET is typically performed on the skin. The body surface dose can be non-uniform in VMAT techniques thus reducing the accuracy of in vivo dosimetry. This may be one reason why centres choose not to perform in vivo dosimetry.
With a move to more optimised conformal TBI techniques with complex deliveries, recommendations for appropriate methods of verification and suitable tolerances become increasingly important. A final safety check should also include an external, independent dosimetry audit.
Conclusion
Optimised conformal TBI techniques offer the ability to reduce doses to OARs, which could lead to a reduction in toxicity. These advanced techniques, combined with organ and target delineation, will also allow for reporting on a wider range of plan dosimetry parameters enabling more in-depth research into correlation with clinical outcomes and toxicity. However, in order to do so and for these techniques to be appropriately incorporated into future clinical trials, there must be standardisation and collaboration within the international TBI community. The key finding of this review is the limited yet varied extent to which TBI planning, delivery and dosimetry protocols are reported in the literature. This alone demonstrates the need for standardisation, e.g. following the ESTRO RATINGS methodology 58 for reporting radiotherapy planning techniques. Consensus guidance is required for PTV outlining, setting planning objectives and OAR dose-volume constraints and recommendations provided on verification and dosimetric audit of these complex techniques to confirm delivery accuracy. Only then can we fully evaluate and research the potential benefits of these optimised conformal methods for TBI treatment.
Contributor Information
Sarah Misson-Yates, Email: sarah.misson-yates@kcl.ac.uk.
Rissa Cunningham, Email: rissa.cunningham@gstt.nhs.uk.
Regina Gonzalez, Email: regina.gonzalez@gstt.nhs.uk.
Patricia Diez, Email: patricia.diez@nhs.net.
Catharine H Clark, Email: catharine.clark@nhs.net.
REFERENCES
- 1.Wong JYC, Filippi AR, Dabaja BS, Yahalom J, Specht L. Total body irradiation: guidelines from the international lymphoma radiation oncology group (ILROG). Int J Radiat Oncol Biol Phys 2018; 101: 521–29: S0360-3016(18)30749-1. doi: 10.1016/j.ijrobp.2018.04.071 [DOI] [PubMed] [Google Scholar]
- 2.Hoeben BAW, Pazos M, Seravalli E, Bosman ME, Losert C, Albert MH, et al. ESTRO ACROP and SIOPE recommendations for myeloablative total body irradiation in children. Radiother Oncol 2022; 173: 119–33: S0167-8140(22)04133-0. doi: 10.1016/j.radonc.2022.05.027 [DOI] [PubMed] [Google Scholar]
- 3.Vogel, J, Hui S, Hua CH, Dusenbery K, Rassiah P, Kalapurakal J, et al. Pulmonary toxicity after total body irradiation. Critical Review of the Literature and Recommendations for Toxicity Reporting Frontiers in Oncology 2021; 11: 708906. doi: 10.3389/fonc.2021.708906 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Wolden, SL, Rabinovitch RA, Bittner NH, Galvin JM, Giap HB, Schomberg PJ, et al. American college of radiology (ACR) and american society for radiation oncology (ASTRO) practice guideline for the performance of total body irradiation (TBI. American Journal of Clinical Oncology 2013; 36: 97–101. Available from: 10.1097/COC.0b013e31826e0528 [DOI] [PubMed] [Google Scholar]
- 5.Van Dyk J, Galvin, JM, Glasgow GP, Podgorsak EB. American association of physicists in medicine report 17 of task group 29 the physical aspects of total and half body irradiation. Association of Physicists in Medicine 1986. Available from: https://doi.org/10.37206/16 [Google Scholar]
- 6.Tas B, Durmus IF, Okumus A, Uzel OE, Gokce M, Goksoy HS, et al. Total-Body irradiation using linac-based volumetric modulated Arc therapy: its clinical accuracy, feasibility and reliability. Radiother Oncol 2018; 129: 527–33: S0167-8140(18)33443-1. doi: 10.1016/j.radonc.2018.08.005 [DOI] [PubMed] [Google Scholar]
- 7.Losert C, Shpani R, Kießling R, Freislederer P, Li M, Walter F, et al. Novel rotatable tabletop for total-body irradiation using a linac-based VMAT technique. Radiat Oncol 2019; 14: 244: 244. doi: 10.1186/s13014-019-1445-3 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Yaray K, Damulira E. Evaluation of volumetric modulated Arc therapy (VMAT)-based total body irradiation (TBI) in pediatric patients. Rep Pract Oncol Radiother 2021; 26: 518–27. doi: 10.5603/RPOR.a2021.0061 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Springer A, Hammer J, Winkler E, Track C, Huppert R, Böhm A, et al. Total body irradiation with volumetric modulated arc therapy: dosimetric data and first clinical experience. Radiat Oncol 2016; 11: 46. doi: 10.1186/s13014-016-0625-7 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Broggi S, Fiorino C, Chiara A, Salvadori G, Peccatori J, Assanelli A, et al. Clinical implementation of low-dose total body irradiation using topotherapy technique. Phys Imaging Radiat Oncol 2019; 12: 74–79. doi: 10.1016/j.phro.2019.11.009 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Cleuziou J-P, Desgranges C, Henry I, Jaumot M, Chartier P, Sihanath R, et al. Total body irradiation using helical tomotherapy: set-up experience and in-vivo dosimetric evaluation. Cancer Radiother 2021; 25: 213–21: S1278-3218(20)30339-5. doi: 10.1016/j.canrad.2020.07.009 [DOI] [PubMed] [Google Scholar]
- 12.Konishi T, Ogawa H, Najima Y, Hashimoto S, Wada A, Adachi H, et al. Safety of total body irradiation using intensity-modulated radiation therapy by helical tomotherapy in allogeneic hematopoietic stem cell transplantation: a prospective pilot study. J Radiat Res 2020; 61: 969–76. doi: 10.1093/jrr/rraa078 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Sun R, Cuenca X, Itti R, Nguyen Quoc S, Vernant JP, Mazeron JJ, et al. First French experiences of total body irradiations using helical tomotherapy® Cancer Radiother 2017; 21: 365–72: S1278-3218(17)30093-8. doi: 10.1016/j.canrad.2017.01.014 [DOI] [PubMed] [Google Scholar]
- 14.Loginova AA, Tovmasian DA, Lisovskaya AO, Kobyzeva DA, Maschan MA, Chernyaev AP, et al. Optimized conformal total body irradiation methods with helical tomotherapy and elekta VMAT: implementation, imaging, planning and dose delivery for pediatric patients. Front Oncol 2022; 12: 785917. doi: 10.3389/fonc.2022.785917 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Rassiah P, Esiashvili N, Olch AJ, Hua C-H, Ulin K, Molineu A, et al. Practice patterns of pediatric total body irradiation techniques: a children’s Oncology Group survey. Int J Radiat Oncol Biol Phys 2021; 111: 1155–64: S0360-3016(21)02628-6. doi: 10.1016/j.ijrobp.2021.07.1715 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Hoeben BAW, Pazos M, Albert MH, Seravalli E, Bosman ME, Losert C, et al. Towards homogenization of total body irradiation practices in pediatric patients across SIOPE affiliated centers. A survey by the SIOPE radiation oncology Working group. Radiother Oncol 2021; 155: 113–19: S0167-8140(20)30879-3. doi: 10.1016/j.radonc.2020.10.032 [DOI] [PubMed] [Google Scholar]
- 17.Peters M, Taylor B, Turner E. An evidence-based review of total body irradiation. J Med Imaging Radiat Sci 2015; 46: 442–49: S1939-8654(15)00272-6. doi: 10.1016/j.jmir.2015.09.007 [DOI] [PubMed] [Google Scholar]
- 18.Paix A, Antoni D, Waissi W, Ledoux MP, Bilger K, Fornecker L, et al. Total body irradiation in allogeneic bone marrow transplantation conditioning regimens: a review. Crit Rev Oncol Hematol 2018; 123: 138–48: S1040-8428(17)30326-8. doi: 10.1016/j.critrevonc.2018.01.011 [DOI] [PubMed] [Google Scholar]
- 19.Hill-Kayser CE, Plastaras JP, Tochner Z, Glatstein E. Tbi during BM and SCT: review of the past, discussion of the present and consideration of future directions. Bone Marrow Transplant 2011; 46: 475–84. doi: 10.1038/bmt.2010.280 [DOI] [PubMed] [Google Scholar]
- 20.Wong JYC, Filippi AR, Scorsetti M, Hui S, Muren LP, Mancosu P. Total marrow and total lymphoid irradiation in bone marrow transplantation for acute leukaemia. Lancet Oncol 2020; 21: e477–87: S1470-2045(20)30342-9. doi: 10.1016/S1470-2045(20)30342-9 [DOI] [PubMed] [Google Scholar]
- 21.Thomas DH, Miller B, Rabinovitch R, Milgrom S, Kavanagh B, Diot Q, et al. Integration of automation into an existing clinical workflow to improve efficiency and reduce errors in the manual treatment planning process for total body irradiation (TBI). J Appl Clin Med Phys 2020; 21: 100–106. doi: 10.1002/acm2.12894 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Teruel JR, Taneja S, Galavis PE, Osterman KS, McCarthy A, Malin M, et al. Automatic treatment planning for VMAT-based total body irradiation using eclipse scripting. J Appl Clin Med Phys 2021; 22: 119–30. doi: 10.1002/acm2.13189 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Frederick R, Hudson A, Balogh A, Cao JQ, Pierce G. Standardized flattening filter free volumetric modulated Arc therapy plans based on anteroposterior width for total body irradiation. J Appl Clin Med Phys 2020; 21: 75–86. doi: 10.1002/acm2.12827 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Studinski RCN, Fraser DJ, Samant RS, MacPherson MS. Current practice in total-body irradiation: results of a Canada-wide survey. Curr Oncol 2017; 24: 181–86. doi: 10.3747/co.24.3484 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Fog LS, Wirth A, MacManus M, Downes S, Grace M, Moggré A, et al. Total body irradiation in Australia and New Zealand: results of a practice survey. Phys Eng Sci Med 2020; 43: 825–35. doi: 10.1007/s13246-020-00878-z [DOI] [PubMed] [Google Scholar]
- 26.Koken PW, Murrer LHP. Total body irradiation and total skin irradiation techniques in Belgium and the Netherlands: current clinical practice. Adv Radiat Oncol 2021; 6: 100664: 100664. doi: 10.1016/j.adro.2021.100664 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Ishibashi N, Soejima T, Kawaguchi H, Akiba T, Hasegawa M, Isobe K, et al. National survey of myeloablative total body irradiation prior to hematopoietic stem cell transplantation in Japan: survey of the Japanese radiation oncology Study Group (JROSG). J Radiat Res 2018; 59: 477–83. doi: 10.1093/jrr/rry017 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Giebel S, Miszczyk L, Slosarek K, Moukhtari L, Ciceri F, Esteve J, et al. Extreme heterogeneity of myeloablative total body irradiation techniques in clinical practice: a survey of the acute leukemia Working Party of the European group for blood and marrow transplantation. Cancer 2014; 120: 2760–65. doi: 10.1002/cncr.28768 [DOI] [PubMed] [Google Scholar]
- 29.Eaton DJ, Bolton S, Thomas RAS, Clark CH. Inter-departmental dosimetry audits-development of methods and lessons learned. J Med Phys 2015; 40: 183–89. doi: 10.4103/0971-6203.170791 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Clark CH, Aird EGA, Bolton S, Miles EA, Nisbet A, Snaith JAD, et al. Radiotherapy dosimetry audit: three decades of improving standards and accuracy in UK clinical practice and trials. Br J Radiol 2015; 88(1055): 20150251. doi: 10.1259/bjr.20150251 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Izewska J, Bokulic T, Kazantsev P, Wesolowska P, van der Merwe D. 50 years of the IAEA/WHO postal dose audit programme for radiotherapy: what can we learn from 13756 results? Acta Oncol 2020; 59: 495–502. doi: 10.1080/0284186X.2020.1723162 [DOI] [PubMed] [Google Scholar]
- 32.Clark CH, Jornet N, Muren LP. The role of dosimetry audit in achieving high quality radiotherapy. Phys Imaging Radiat Oncol 2018; 5: 85–87. doi: 10.1016/j.phro.2018.03.009 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Izewska J, Lechner W, Wesolowska P. Global availability of dosimetry audits in radiotherapy: the IAEA dosimetry audit networks database. Phys Imaging Radiat Oncol 2018; 5: 1–4. doi: 10.1016/j.phro.2017.12.002 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Lye J, Kry S, Shaw M, Gibbons F, Keehan S, Lehmann J, et al. A comparison of IROC and ACDS on-site audits of reference and non-reference dosimetry. Med Phys 2019; 46: 5878–87. doi: 10.1002/mp.13800 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.NHS England . External Beam Service Specifications. Online]. 2019. Available from: https://www.england.nhs.uk/wp-content/uploads/2019/01/External-Beam-Radiotherapy-Services-Delivered-as-Part-of-a-Radiotherapy-Network-Adults.pdf
- 36.Meeting AV . 07/12/20; 301539; BReP-SNAP-T-92 Topic: Photon External Beam Therapy Accessed from. Available from: https://virtual.aapm.org/aapm/2020/eposters/301539/andrea.molineu.iroc.s.tbi.phantom.results.html?f=listing%3D3%2Abrowseby%3D8%2Asortby%3D2%2Amedia%3D2
- 37.Nelligan, R, Bailey M, Tran T, Baldwin Z. ACPSEM ROSG TBI working group recommendations for quality assurance in total body irradiation. Australasian Physical & Engineering Sciences in Medicine 2015; 38: 205–15. Available from: 10.1007/s13246-015-0344-7 [DOI] [PubMed] [Google Scholar]
- 38.Commissie, N, Stralingsdosimetrie V. Code of Practice and recommendations for Total Body Irradiation and Total Skin Irradiation. February. 2021. Available from: 10.25030/ncs-034 [DOI]
- 39.Dibs K, Sim AJ, Peñagaricano JA, Latifi K, Garcia GA, Peters JA, et al. Gonadal-sparing total body irradiation with the use of helical tomotherapy for nonmalignant indications. Rep Pract Oncol Radiother 2021; 26: 153–58. doi: 10.5603/RPOR.a2021.0006 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40.Isobe A, Usui K, Hara N, Sasai K. The effects of rotational setup errors in total body irradiation using helical tomotherapy. J Appl Clin Med Phys 2021; 22: 93–102. doi: 10.1002/acm2.13271 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41.Kasai Y, Fukuyama Y, Terashima H, Nakamura K, Sasaki T. Dose evaluation indices for total body irradiation using tomodirect with different numbers of ports: a comparison with the tomohelical method. J Appl Clin Med Phys 2019; 20: 129–35. doi: 10.1002/acm2.12540 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42.Salz H, Bohrisch B, Howitz S, Banz N, Weibert K, Wiezorek T, et al. Intensity-modulated total body irradiation (TBI) with tomodirect. Radiat Oncol 2015; 10: 58: 58. doi: 10.1186/s13014-015-0362-3 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43.Takenaka R, Haga A, Yamashita H, Nakagawa K. Adequate target volume in total-body irradiation by intensity-modulated radiation therapy using helical tomotherapy: a simulation study. J Radiat Res 2017; 58: 210–16. doi: 10.1093/jrr/rrw115 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44.Sresty NVNM, Gudipudi D, Krishnam Raju A, Anil kumar T, Lakshmi VRP, Srikanth G, et al. Total body irradiation of bone marrow transplant using helical tomotherapy with a focus on the quality of dose contribution at junction target volumes. Strahlenther Onkol 2021; 197: 722–29. doi: 10.1007/s00066-021-01769-2 [DOI] [PubMed] [Google Scholar]
- 45.Lin Y, Kong F, Li H, Xu D, Jia F, Zhang X, et al. Comparison of target volume and clinical effects of four radiotherapy plans for acute lymphoblastic leukemia prior to hematopoietic stem cell transplantation. Mol Med Report 2018; 18: 2762–70. doi: 10.3892/mmr.2018.9228 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 46.Morrison CT, Symons KL, Woodings SJ, House MJ. Verification of junction dose between VMAT arcs of total body irradiation using a sun nuclear arccheck phantom. J Appl Clin Med Phys 2017; 18: 177–82. doi: 10.1002/acm2.12208 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 47.Stanley, D, McConnell K, Iqbal Z, Everett A, Dodson J, Keene K, et al. Dosimetric evaluation between the conventional volumetrically modulated arc therapy (VMAT). Total Body Irradiation (TBI) and the Novel Simultaneous Integrated Total Marrow Approach (SIMBa) VMAT TBI Cureus 2021; 13: e15646. doi: 10.7759/cureus.15646 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 48.Pierce G, Balogh A, Frederick R, Gordon D, Yarschenko A, Hudson A. Extended SSD VMAT treatment for total body irradiation. J Appl Clin Med Phys 2019; 20: 200–211. doi: 10.1002/acm2.12519 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 49.Zhang-Velten ER, Parsons D, Lee P, Chambers E, Abdulrahman R, Desai NB, et al. Volumetric modulated Arc therapy enabled total body irradiation (VMAT-TBI): six-year clinical experience and treatment outcomes. Transplant Cell Ther 2022; 28: 113. S2666-6367(21)01359-2. doi: 10.1016/j.jtct.2021.10.020 [DOI] [PubMed] [Google Scholar]
- 50.Guo B, Sheen C, Murphy E, Magnelli A, Lu L, Cho Y, et al. Image-Guided volumetric-modulated Arc therapy of total body irradiation: an efficient workflow from simulation to delivery. J Appl Clin Med Phys 2021; 22: 169–77. doi: 10.1002/acm2.13412 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 51.Kovalchuk N, Simiele E, Skinner L, Yang Y, Howell N, Lewis J, et al. The Stanford volumetric modulated Arc therapy total body irradiation technique. Pract Radiat Oncol 2022; 12: 245–58: S1879-8500(21)00350-7. doi: 10.1016/j.prro.2021.12.007 [DOI] [PubMed] [Google Scholar]
- 52.Durie E, Nicholson E, Anthias C, Dunne EM, Potter M, Ethell M, et al. Determining the incidence of interstitial pneumonitis and chronic kidney disease following full intensity haemopoetic stem cell transplant conditioned using a forward-planned intensity modulated total body irradiation technique. Radiother Oncol 2021; 158: 97–103: S0167-8140(21)06083-7. doi: 10.1016/j.radonc.2021.02.020 [DOI] [PubMed] [Google Scholar]
- 53.Fog LS, Hansen VN, Kjær-Kristoffersen F, Berlon TE, Petersen PM, Mandeville H, et al. A step and shoot intensity modulated technique for total body irradiation. Tech Innov Patient Support Radiat Oncol 2019; 10: 1–7. doi: 10.1016/j.tipsro.2019.05.002 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 54.van Leeuwen RGH, Verwegen D, van Kollenburg PGM, Swinkels M, van der Maazen RWM. Early clinical experience with a total body irradiation technique using field-in-field beams and on-line image guidance. Phys Imaging Radiat Oncol 2020; 16: 12–17. doi: 10.1016/j.phro.2020.09.004 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 55.Hansen AT, Rose HK, Yates ES, Hansen J, Petersen JBB. Two compound techniques for total body irradiation. Tech Innov Patient Support Radiat Oncol 2022; 21: 1–7. doi: 10.1016/j.tipsro.2021.11.006 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 56.Effeney B, Biggs J, Brady C, Pemberton M, Sim L, Pullar A. Considerations and adaptions to the modulated Arc total body irradiation technique: dosimetry description. J Med Radiat Sci 2019; 66: 284–91. doi: 10.1002/jmrs.363 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 57.Härtl PM, Treutwein M, Hautmann MG, März M, Pohl F, Kölbl O, et al. Total body irradiation-an attachment free sweeping beam technique. Radiat Oncol 2016; 11: 81: 81. doi: 10.1186/s13014-016-0658-y [DOI] [PMC free article] [PubMed] [Google Scholar]
- 58.Hansen CR, Crijns W, Hussein M, Rossi L, Gallego P, Verbakel W, et al. Radiotherapy treatment planning study guidelines (rating): a framework for setting up and reporting on scientific treatment planning studies. Radiother Oncol 2020; 153: 67–78: S0167-8140(20)30808-2. doi: 10.1016/j.radonc.2020.09.033 [DOI] [PubMed] [Google Scholar]
- 59.Hodapp N. Der ICRU-report 83: verordnung, dokumentation und kommunikation Der fluenzmodulierten photonenstrahlentherapie (IMRT). [The ICRU Report 83]. Strahlenther Onkol 2012; 188: 97–100. doi: 10.1007/s00066-011-0015-x [DOI] [PubMed] [Google Scholar]
- 60.Hoeben BAW, Wong JYC, Fog LS, Losert C, Filippi AR, Bentzen SM, et al. Total body irradiation in haematopoietic stem cell transplantation for paediatric acute lymphoblastic leukaemia: review of the literature and future directions. Front Pediatr 2021; 9: 774348: 774348. doi: 10.3389/fped.2021.774348 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 61.Wills C, Cherlon S, Yousef J, Wang K, Mackley H.. Total body irradiation: a practical review. Appl Rad Oncol 2016;5:11–7 [Google Scholar]
- 62.Sabloff M, Tisseverasinghe S, Babadagli ME, Samant R. Total body irradiation for hematopoietic stem cell transplantation: what can we agree on? Curr Oncol 2021; 28: 903–17. doi: 10.3390/curroncol28010089 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 63.Cutter DJ, Ramroth J, Diez P, Buckle A, Ntentas G, Popova B, et al. Predicted risks of cardiovascular disease following chemotherapy and radiotherapy in the UK NCRI rapid trial of positron emission tomography-directed therapy for early-stage Hodgkin lymphoma. J Clin Oncol 2021; 39: 3591–3601. doi: 10.1200/JCO.21.00408 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 64.Diesch-Furlanetto T, Gabriel M, Zajac-Spychala O, Cattoni A, Hoeben BAW, Balduzzi A. Late effects after haematopoietic stem cell transplantation in all, long-term follow-up and transition: a step into adult life. Front Pediatr 2021; 9: 773895: 773895. doi: 10.3389/fped.2021.773895 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 65.Glenn MC, Wallner K, Luk SMH, Ermoian R, Tseng YD, Phillips M, et al. Impact of lung block shape on cardiac dose for total body irradiation. Phys Imaging Radiat Oncol 2022; 21: 30–34. doi: 10.1016/j.phro.2022.01.004 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 66.Hussein M, Clark CH, Nisbet A. Challenges in calculation of the gamma index in radiotherapy-towards good practice. Phys Med 2017; 36: 1–11: S1120-1797(17)30055-8. doi: 10.1016/j.ejmp.2017.03.001 [DOI] [PubMed] [Google Scholar]
- 67.Miften M, Olch A, Mihailidis D, Moran J, Pawlicki T, Molineu A, et al. Tolerance limits and methodologies for IMRT measurement-based verification QA: recommendations of AAPM task group No. 218. Med Phys 2018; 45: e53–83. doi: 10.1002/mp.12810 [DOI] [PubMed] [Google Scholar]