Skip to main content
. 2023 Apr 7;881:163292. doi: 10.1016/j.scitotenv.2023.163292

Table 3.

Performance metrics of the N200 VOC assay for each of the probes in singleplex, duplex or triplex.

Assay Slope y-intercept Efficiency R2
Universal Singleplex −3.347 40.236 98.9 0.990
Duplex (with R203M) −3.366 39.50 98.2 0.991
Duplex (with R203K + G204R) −3.299 38.793 101.0 0.990
Duplex (with S202S) −3.311 40.207 100.5 0.990
Duplex (with T205I) −3.461 39.773 94.5 0.990
Triplex (with R203M, R203K + G204R) −3.443 39.661 95.2 0.995
R203K + G204R
(Omicron/Alpha)
Singleplex −3.429 38.834 95.7 0.998
Duplex (with Universal) −3.334 39.585 99.5 0.991
Triplex (with Universal, R20K + G204R) −3.464 39.902 94.4 0.995
T205I
(Beta)
Singleplex −3.396 39.951 97.0 0.997
Duplex (with Universal) −3.533 39.952 91.9 0.994
S202S
(Gamma)
Singleplex −3.249 40.363 103.1 0.998
Duplex (with Universal) −3.352 39.891 98.8 0.988
R203M
(Delta)
Singleplex −3.541 40.365 91.6 0.998
Duplex (with Universal) −3.387 39.513 97.4 0.990
Triplex (with Universal, R203M) −3.442 39.124 95.2 0.995