Table 2. Definitions of Assay Parameters34.
LoB | calculated concentration based on the mean zero calibrator signal + 1.645 × the standard deviation (SD) of the zero calibrator |
LLoD | calculated concentration based on the signal + 1.645 × SD (lowest calibrator) above the LoB |
LLoQ | calculated concentration based on the signal + 10 × SD (blank) above the zero calibrator |
ULoQ | calibrator concentration at the upper limit of the linear range on a logarithmic scale |