Examples of scintillation materials used in PET detectors (data from [11], [12]). Zeff is the effective atomic number.

Material Density (g/cm3) Zeff Light output (photons/keV) Decay time (ns) Hygroscopic
NaI:Tl 3.67 51 41 230 yes
BGO 7.1 75 9 300 no
L(Y)SO:Ce 7.1-7.4 65-66 26-34 38-44 no
BaF2 4.9 54 1.3-1.4 0.8 no
CsF 4.6 52 1.9-2.0 3 yes
LaBr3:Ce 5.1 45 64-76 16 yes