TABLE 1.

Overview of the reported specifications for various RS-OCT system configurations

Multimodal integration strategy Author Excitation Laser power at sample (RS) Acquisition time (RS) Spectral resolution (RS) Spot size (RS) Spatial resolution (OCT)
Axial Lateral
Shared RS–OCT sample arm, separate detection systems Patil et al. [76] RS: 785 nm, FP region
TD-OCT: 1310 ± 30 nm
40 mW 7.7 cm−1 75 μm 21 μm 25 μm
Evans et al. [75] RS: 633 nm, FP region
SD-OCT: 855 ± 37 nm
4 mW 5 s ~6 cm−1 15 mm 4.5 μm 18 μm
Patil et al. [77] RS: 785 nm, FP region
TD-OCT: 1310 ± 30 nm
40 mW 30 s 44 μm 14 μm 25 μm
Khan et al. [82] RS: 785 nm, FP region
TD-OCT: 1310 ± 21 nm
60 mW 10 s 8 cm−1 12 μm 20 μm 15 μm
Maher et al. [83] RS: 785 nm, FP region
SD-OCT: 840 ± 22 nm
30 mW 3.3 cm−1 8 μm 40 μm
Zakharov et al. [29] RS: 785 nm, FP region
SD-OCT: 840 ± 50 nm
30 s 0.05 nm 1 mm 3.5 μm 3.5 μm
Wang et al. [78] RS: 785 nm, FP/HW regions
SS-OCT: 1325 ± 60 nm
10–15 mW 0.5 s ~9 cm−1 12 μm 12 μm
Chen et al. [91] RS: 785 nm, FP region
SD-OCT: 850 ± 15 nm
100 mW 2 s 15 μm
Mazurenka et al. [80] RS: 532 nm pulsed, FP region
SD-OCT: 1325 nm
2.38 mJ/cm2 100 s Collimated beam 5.5 μm 13 μm
Ren et al. [84] RS: 532 nm, FP region
SD-OCT: 830 ± 20 nm
10 mW 30–60 s 7 cm−1 3 μm 2.3 μm 7.35 μm
Shared RS-OCT sample arm and detection system Patil, et al. [73] RS: 785 nm, FP region
SD-OCT: 855 ± 20 nm
100 mW 30–90 s 7 cm−1 15 μm 8 μm 18 μm
Separate RS-OCT sample arms, manual or automated coregistration Klemes, et al. [79] RS: 632.5 nm, FP region
SD-OCT: 880 ± 32 nm
13 mW 20–50 s 4.3 μm 3.6 μm
Placzek, et al. [86] RS: 785 nm, FP region
SS-OCT: 1304 ± 45 nm
70 mW 5 s 5 cm−1 100 μm 12 μm 28 μm

Abbreviations: FP, fingerprint; HW, high wavenumber; CT, optical coherence tomography; RS: Raman Spectroscopy; SD, spectral domain; SS, swept source; TD, time domain.