. 2023 Mar 5;13(5):1594–1606. doi: 10.7150/thno.81938

Table 1.

Multiparametric MRI sequence parameters.

Sequence T2w TSE STIR T1w TSE Dixon T2w TSE IVIM DWI VIBE
DCE
T1w TSE Dixon T1w VIBE
Dixon
Additional features 2D 2D 2D 2D 2D 3D 2D 3D
Plane Cor Cor Tra Tra Tra Tra Cor Tra
Echo time (TE; ms) 40 10 104 64 61 1.14 9.8 2.46, 3.69
Repetition time (TR; ms) 5600 650 11900 6900 7780 3.02 650 5.5
Field of view (FOV; mm) 500 500 220 220 220 220 500 350
Slice thickness (mm) 4 4 4 4 4 4 4 3
In-plane resolution (mm2) 1.0 × 1.0 0.8 × 0.8 0.3 × 0.3 1.8 × 1.8 0.9 x 0.9 1.1 x 1.1 0.9 x 0.9 0.9 x 0.9
Flip angle (°) 140 160 160 90 180 15 90 9
Number of slices 30 30 40 40 40 40 30 112
Slice distance (mm) 4.8 4.8 4 4 4 4 4.8 3
Echo Train Length 12 3 25 53 0 1 3 2
Bandwidth per pixel (Hz) 222 845 200 1812 883 789 868 690
Phase Encoding Direction Row Row Col Col Col Col Row Col
Number of averages 1 1 3 3 1 1 2 3
Acquisition time (min) 01:54 01:36 03:48 04:45 08:43 04:33 04:24 03:08
Fat water contrast Standard Fast Dixon Standard SPAIR Fat Saturation Standard Fast Dixon Dixon
Fat Saturation STIR Strong - Strong Strong - Strong -
Diffusion Mode - - - 3D Diagonal 4-Scan Trace - - -
B value (s/mm2) - - - 0, 10, 20, 30, 40, 50, 70, 100, 200, 400, 600, 800 50, 400, 1000 - - -
Calculated b value (s/mm2) - - - - 1400 - - -
EPI factor - - - 90 95 - - -
Contrast enhancement No No No No No Yes Yes Yes
Temporal resolution (s) - - - - - 5.5 - -
Measurements - - - - - 50 - -

Col: column; Cor: coronal; DCE: dynamic contrast enhanced imaging; DWI: diffusion weighted imaging; EPI: echo-planar imaging; FOV: field of view; fs: fat-saturated; IVIM: intravoxel incoherent motion; SPAIR: spectral attenuated inversion recovery; STIR: short tau inversion recovery; Tra: transverse; VIBE: volumetric interpolated breath-hold examination; w: weighted