Skip to main content
. 2022 Nov 24;50(1):480–494. doi: 10.1002/mp.16095

TABLE 2.

Summary of the linear attenuation coefficient (LAC: μ) for multiple tissue substitutes39

Lung Water Griffith‐Bone

Threshold

in this study

Modality

Tube voltage

(kV)

E eff

(keV)

μ

(cm‐1)

μ scat

(cm‐1)

μ

(cm‐1)

μ scat

(cm‐1)

μ

(cm‐1)

μ scat

(cm‐1)

μ 200

(cm‐1)

XF 50 28 0.114 0.065 0.399 0.235 1.150 0.304 0.642
XF 70 31 0.098 0.063 0.345 0.227 0.915 0.293 0.529
XF 90 34 0.088 0.062 0.309 0.222 0.753 0.283 0.453
CT 120 56 0.059 0.052 0.212 0.196 0.343 0.244 0.254
Calculated CT value (HU) −722 0 618 200

Note: Lung, Water and Griffith‐Bone are tissue substitutes.39

Abbreviations: CT, computer tomography; E eff, effective energy; HU, Hounsfield unit; μ scat, scatter contribution of LAC; μ 200, LAC threshold to separate contribution of soft tissue from that of bone; XF, X‐ray fluoroscopy.