Skip to main content
. 2023 Mar 1;13(4):2262–2277. doi: 10.21037/qims-22-596

Figure 3.

Figure 3

Linear regression analysis and Bland-Altman plots between RTCSCineMoCo and SegBH (A-J) and between RTCSCine and SegBH (K-T) for LV functional parameters including LVEF, LVEDV, LVESV, LVSV, and LVM. (—) Mean difference; (---) 95% CI limits of agreement. CI, confidence interval; LV, left ventricular; LVEDV, left ventricular end-diastolic volume; LVEF, left ventricular ejection fraction; LVESV, left ventricular end-systolic volume; LVM, left ventricular mass; LVSV, left ventricular stroke volume; r, correlation coefficient; RTCSCine, real-time cine imaging with compressed sensing; RTCSCineMoCo, real-time cine imaging combines compressed sensing reconstruction and retrospective fully automated respiratory motion correction; SegBH, segmented acquisition with retrospective electrocardiogram gating and breath-hold.