Skip to main content
. 2023 Mar 13;18(5):1924–1928. doi: 10.1016/j.radcr.2023.02.013

Fig. 2.

Fig 2

Lateral skull x-ray completed post 3T-MRI demonstrating a right frontal approach ventriculoperitoneal shunt (Codman-Hakim Programmable valve) with a displaced pressure control cam (black arrow) from the now apparent X-shaped base plate of the plastic valve housing.