Table 1.

Cases of prior and current pressure control cam displacements reported in patients with the Codman-Hakim programmable valve shunt system.

Author Age (years)/Gender at presentation Time from placement to symptoms Mechanism of displacement Symptoms Time from mechanism to symptoms Ventricular size at presentation Valve replacement Crack atop plastic housing Valve inspection findings Postop radiographic findings Outcome
Okazaki et al. [9] 7 (M) 7 Blow behind right ear over valve Positional headache 2 days Slit like ventricles Yes Yes Pressure control cam out of plastic valve housing None reported Resolution of symptoms
Watanabe et al. [8] 57 (F) 8 Valve adjustment Worsening of chronic headache and nausea Few days Slit like ventricles Yes Yes Pressure control cam displaced distally CTH showing resolution of slit ventricles
Kim et al. [7] 72 (F) 9 Completion of 3T MRI Postural headache, neck pain, nausea Several hours Decreased ventricular size Yes Yes Pressure control cam displaced distally CTH showing resolution of decreased ventricle size
Castillo et al. 8 (F) 8 Completion of 3T MRI None N/A Not preformed Yes, to MPV No Not reported Not performed

3T MRI, 3-Tesla magnetic resonance imaging; CTH, commuted tomography of the head; F, female; M, male; MPV, medium pressure valve; N/A, not applicable.