Skip to main content
. 2022 Dec 20;392(1):301–306. doi: 10.1007/s00441-022-03715-9

Table 1.

Overview on the diagnostic accuracy of the CSF RT-QuIC in different studies

Species Amino acid res Accuracy for sCJD CSF diagnostics
Sensitivity / Specificity
References
Human 23–231  > 80% (27/34) 100% (0/49) Atarashi et al. 2011
Hamster 23–231

89% (109/123)

88.2% (148/179)

99% (1/103)

99.4% (1/163)

McGuire et al. 2012

Lattanzio et al. 2017

90–231

94% (106/113)

92% (102/111)

95% (60/63)

100% (0/64)

98.5% (1/67)

100% (14/14)

Orrú et al. 2015a; Groveman et al. 2016

Foutz et al. 2017

Foutz et al. 2017

 90-231 96% (58/61) 100% (0/80) Fiorini et al. 2020
Bank vole  23-230 88.6% (70/79) 91.2% (5/57) Orrú et al. 2015b; Mok et al. 2021
Hamster-sheep chimera

Hamster

23–137

Sheep 141–234

80% (51/64)

89% (163/183)

99% (2/400)

100% (0/118)

Cramm et al. 20152016

Hermann et al. 2018