Table 1.

MRI protocol for scanning knee arthroplasty at 1.5T

Imaging Series
Variable 2D FSE 3D Iso-MAVRIC MAVRIC T2 MAPPING
FOV Axial, Coronal Sagittal Sagittal Sagittal
In-plane Frequency Direction Ax: S to I
Cor: R to L
Sag: A to P
A to P S to I
Receiver Bandwidth (kHz) ±125 ±125 ±125
Flip Angle (°) 90 110 100
Repetition Time (ms) 4505 3500 3500
Echo Time (ms) 27 8.8 TE1: 9.8; TE2: 90
Echo Train length 16 48 40
Acquisition Matrix (mm) 512 x (256–320) 224 x 224 256 x 192
Field of View (cm) 20 20–22 21–22
Slice Thickness (mm) 3–3.5 1.2 3.5