Skip to main content
. Author manuscript; available in PMC: 2023 Apr 25.
Published in final edited form as: Proc SPIE Int Soc Opt Eng. 2023 Mar 6;12371:1237108. doi: 10.1117/12.2652588

Table 2.

Central axis surface dose for various tested field sizes.

Field Size Halcyon (6 MV FFF) TrueBeam (6 MV) % Difference
5 × 5 cm 42.68 cGy 32.27 cGy 32.2%
10 × 10 cm 51.32 cGy 38.03 cGy 34.9%
20 × 20 cm 57.03 cGy 46.95 cGy 21.5%
28 × 28 cm 60.99 cGy 55.09 cGy 10.7%