Table 5.

The diagnostic accuracy of BI-RADS with DM, BI-RADS with DBT, and BI-RADS with integrated DBT and DM for confident diagnosis of breast malignancy.

Criterion DM DBT DM + DBT
% 95%CI % 95%CI % 95%CI
Accuracy 67.3 90 97.5
Specificity 67.6 60.8–74 90.3 85.5–94 96.6 93.2–98.6
Sensitivity 66.9 58.7–74.4 89.2 83–93.7 98.7 95.2–99.8
Positive likelihood ratio 2.1 1.7–2.6 9.23 6.1–14.1 29.2 14.1–60.4
Negative likelihood ratio 0.49 0.38–0.36 0.12 0.08–0.19 0.01 0–0.06
PPV 59.6 51.8–67.2 86.8 80.4–91.8 95.4 90.8–98.1
NPV 74.1 67.2–80.2 92.1 87.5–95 99 96.5–99.9

BI-RADS: Breast Imaging Reporting and Data System; AUC: area under curve; PPV: positive predictive value; NPV: negative predictive value; CI: confidence interval.