Skip to main content
. 2023 Jan 30;36(2):175–189. doi: 10.1007/s10334-023-01063-5

Fig. 8.

Fig. 8

GIRF (first-order terms) spectra acquired at 11.7 T without and with lead tube (top row: − 5 to + 5 kHz interval, bottom row: zoom on the 0–3 kHz interval). The lead tube on the Z-axis led to unwanted peaks in the ~ 900–1150 Hz region. Removing the lead tube suppressed the peaks in that region but boosted some peaks at higher frequencies