Skip to main content
. 2023 Apr 18;12(8):2942. doi: 10.3390/jcm12082942

Figure 2.

Figure 2

Autotriggering induced by noise in the expiratory phase of the cycle in a ventilator with electronic trigger (shape method). Observe that it cannot be considered as a double triggering because the effort (red vertical line) clearly ended before the start of the second cycle.