Skip to main content
. 2023 Apr 20;12(8):1722. doi: 10.3390/plants12081722

Table 4.

Determination of total phytosterols in dietary supplements by chromatographic methods.

Method Detector Sample Preparation/Extraction Analytical Methods Target
Compounds
Ref.
GC FID
  • Addition internal standard solution

  • Saponification: KOH (2 h/100 °C)

  • Derivatization: BSTFA, and pyridine

Column: (25m × 0.25mm, 0,33 μm,
Dimethylpolysiloxane stationary phase)
Temperatures (°C):
Detector: 325; Injector: 325
Gas (mL/min): Helium (0.5)
Injection type: Split ratio, 1:40
Injection volume (μL): 1
campesterol
stigmasterol
SIT
total sterols
[118]
GC FID/MS
  • Addition internal standard solution (epicoprostanol)

  • Acid hydrolysis HCl (4M/1 h/100 °C)

  • Lipid extraction: EE: diethyl ether: petroleum ether (50:50/v:v)

  • Saponification: ethanolic KOH (2 M/80 °C/60 min)

  • Unsaponifiable extraction (LLE): EE (3 times)

  • Derivatization: BSTFA, and pyridine

Column: (30 m × 0.32 mm, 0.25 μm, (poly (94%methyl/5% phenyl) silicone))
Temperatures (°C): column 250, injector 290, detector 290
Gas (mL/min): H2 (1)
Injection type Split ratio, 1:25
Injection volume (μL): 1
campesterol
stigmasterol
SIT
[22]
GC FID
  • Alkaline and acid/alkaline protocols

  • Addition internal standard solution (epicoprostanol)

  • Acid hydrolysis: HCl (3N)

  • Saponification: NaOH 2.3 N in MeOH + HCl 3N + NaCl

  • Derivatization: BSTFA, and pyridine

Column: (30 m × 0.32 mm, 0.25 μm) (5%-phenyl)-methylpolysiloxane stationary phase)/(30 m × 0.32 mm, 0.25 μm, (5%-phenyl) (1%-vinyl)-methylpolysiloxane stationary phase)
Temperatures (°C): column 250, Injector 290, detector 290
Gas (mL/min): H2 (1)
Injection type Split ratio, 1:25
Injection volume (μL): 1
17 phytosterols [138]
GC FID/MS
  • Addition internal standard solution (epicholesterol)

  • Saponification: aqueous KOH (1.3 N/85–89 °C/20 min)

  • Unsaponifiable extraction: cyclohexane

  • Derivatization: (BSTFA/TMCS)

Column: (15 m × 0.32-mm,0.25 μm/5% diphenyl-95%-Dimethylpolysiloxane)
Temperatures (°C): column 270, injector 280, detector 300.
Gas (ml/min): helium (0.58)
Injector type: split ratio, 17:1
Injection volume (μL): 0.5
SIT
β-sitosteryl glucoside (BSSG)
[139]
GC FID
  • Saponification: ethanolic KOH (1 M/hot plate/80–90 min)

  • Unsaponifiable extraction: toluene

  • Derivatization: TMCS

  • Addition internal standard solution (5 alfa-cholestane)

Column: (25 m × 0.32 mm, 17 μm/5% phenyl-methylsilicone or methyl silicone gum stationary phase)
Temperatures (°C): Column 190→255, injector: 250, detector: 300
Gas (mL/min): helium 2/15/3; makeup helium 20/hydrogen 35/air 380
Injection volume (μL): 1
Campesterol
Stigmasterol
SIT
[119]
GC FID/MS
  • Addition internal standard solution (cholestanol), chloroform

  • Saponification: methanol KOH (0.3M/80 °C/60 min)

  • Unsaponifiable extraction: hexane (twice)

  • Derivatization: MTSA, TMCS, and pyridine

Column: (60 m × 0.25 mm, 0.25 um/5% phenyl methyl siloxane stationary phase)
Temperatures (°C): column 80→325, detector 230.
Gas (mL/min): helium (1)
Injection volume (μL): 1
campesterol
stigmasterol
SIT
total sterols
[120]
LC APCI–MS
  • Solvent extraction: hexane

  • Saponification with base (ethanolic KOH)/acid hydrolysis

  • Solvent extraction of unsaponifiable material: toluene

1. Column: phenyl (150 mm × 3.9 mm, 3,5 μm)
Mobile phase: 58% acetonitrile, 42% water
Flow rate (mL/min): 1.1 (isocratic)
2. Column: ACE C18 (150 mm × 3.0 mm, 3 μm)
Mobile phase: 90% methanol, 10% water
Flow rate (mL/min): 0.80 (isocratic)
Injection volume (μL): 5 for qualitative, 10 for quantitative measurement
campesterol,
cycloartenol, lupenone, lupeol, SIT, and stigmasterol
(Standard Reference Materials containing saw palmetto)
[39]
HPLC ESI-MS
  • Saponification: ethanolic KOH (70 °C/30 min)

  • Unsaponifiable extraction: hexane

  • Derivatization: dansyl chloride (40 °C/30 min)

Column: C18 (250 × 3.0 mm, 5 μm)
Mobile phase: 95% methanol: 5% water
Flow rate (mL/min): 0.5 (isocratic)
campesterol
stigmasterol
β-sitosterol
[141]
HPLC UV
  • Ultrasound-Assisted Emulsification

  • Microextraction (USAEME):

  • Solubilization: methanol

  • Addition: calcium chloride 2mol/L

  • Fat extraction: hexane (heating for 10 min at 50 °C/ultrasonic time 25 min/centrifugation 2600 rpm, 15 min)

Column: RP-18 (125-4 mm, 5 μm)
Mobile phase: hexane, propan-1-ol (99.5:0.5; v/v)
Flow rate (mL/min): 0.8−1 (isocratic)
Injection volume (μL): 50
UV detection: 212 nm
β-sitosterol [142]
UHPLC UV/CAD
  • Addition internal standard solution (cholesterol)

  • Saponification: ethanolic KOH (80 °C/30 min)

  • Acid hydrolysis: HCl

  • Solvent extraction: hexane (3 min)

Column: (100 × 2.1 mm, 1,7 μm) Phenyl-hexyl
Mobile phase: acetonitrile: water
Flow rate (mL/min): 0.9 (gradient)
Temperature: 60 °C
Inj. vol. (μL): 2
Run time (min): 8.5
ergosterol, brassicasterol,
campesterol, campestanol,
fucosterol, stigmasterol,
stigmastanol, SIT esterified
form
[122]

GC: gas chromatography; FID: flame ionization detection; BSTFA: bis-(trimethylsilyl) trifluoroacetamide; LLE: Liquid-liquid extraction; MS: mass spectrometry; TMCS: trimethylchlorosilane; MTSFA: N- methyl-N-(trimethylsilyl) fluoroacetamide; LC-liquid chromatography; APCI: atmospheric pressure chemical ionization; ESI: ionization by electrospray; UV: ultraviolet; HPLC: high liquid chromatography; UHPLC: ultra-high liquid chromatography; CAD: Charged aerosol detector.