TABLE 5.
Intra-physician, Inter-physician and Physicians compared with algorithm percentage of inconsistences (I, %) and relevant inconsistences (RI, %) in each experiment. For experiments 2 and 3 also expressed relative reduction of inconsistences with respect to experiment 1 expressed in column “% diff”. For intra-physician, the percentage of inconsistences is estimated over the 50 sets of two radiographs of the same patient (150 sets in each experiment), while for the inter-physician is over the 100 radiographs (300 in each experiment).
Intra-physician | Inter-physician | |||||
---|---|---|---|---|---|---|
I (%) | % diff | I (%) | % diff | RI (%) | % diff | |
Experiment 1 | 6.0 | - | 60.3 | - | 13.3 | - |
Experiment 2 | 2.7 | −55.4% | 44.7 | −25.9% | 3.0 | −77.4% |
Experiment 3 | 2.3 | −61.1% | 41.3 | −31.5% | 1.0 | −92.5% |