. 2023 Apr 21;14:1145285. doi: 10.3389/fgene.2023.1145285

TABLE 1.

Analytical performance validation.

Metrics Results
Single Nucleotide Variants (SNV) True Positives = 164 False Negatives = 1
False Positives = 0 True Negatives = 168
Sensitivity 0.99394 (95% CI = 0.9667–0.9998)
Specificity 1.00000 (95% CI = 0.9783–1.0000)
PPV 1.00000
Accuracy 0.99700 (95% CI = 0.9834–0.9999)
Small insertions/deletions * (1–50 bp) True Positives = 137 False Negatives = 3
False Positives = 0 True Negatives = 168
Sensitivity 0.97857 (95% CI = 0.9387–0.9956)
Specificity 1.00000 (95% CI = 0.9783–1.0000)
PPV 1.00000
Accuracy 0.99026 (95% CI = 0.9718–0.9980)
Structural Variants True Positives = 118 False Negatives = 6
False Positives = 0 True Negatives = 168
Sensitivity 0.95833 (95% CI = 0.8977–0.9820)
Specificity 1.00000 (95% CI = 0.9783–1.0000)
PPV 1.00000
Accuracy 0.97945 (95% CI = 0.9558–0.9924)
Mitochondrial Variants ∼ 100% Heteroplasmy True Positives = 70 False Negatives = 0
False Positives = 1 True Negatives = 34
Sensitivity 1.00000 (95% CI = 0.9487–1.0000)
Specificity 0.97143 (95% CI = 0.8508–0.9993)
PPV 0.98592
Accuracy 0.99048 (95% CI = 0.9481 – 0.9998)
a

Small insertion/deletions include insertions, deletions, and combinations of thereof below 50bp.