Table 3. Noise, NPS peaks, and average spatial frequencies for eight reconstruction methods and four dose levels.

50 mA 100 mA 150 mA 200 mA
Noise (HU)
   FBP 41.3043 30.5444 25.0197 21.4577
   AV-30 32.5606 24.0994 19.7209 16.8955
   AV-50 26.7706 19.8267 16.2084 13.8587
   AV-80 18.4297 13.6588 11.1071 9.4756
   AV-100 13.1694 10.0578 8.1843 6.9447
   DL-L 27.3244 19.4344 16.0818 13.9488
   DL-M 21.9028 15.8037 13.1265 11.4632
   DL-H 16.0927 11.8503 9.9109 8.7437
fpeak (mm–1)
   FBP 0.2362 0.2677 0.2992 0.2835
   AV-30 0.2047 0.2577 0.2362 0.2677
   AV-50 0.2047 0.2205 0.2047 0.2520
   AV-80 0.1890 0.1732 0.1260 0.1417
   AV-100 0.1575 0.1102 0.1260 0.1417
   DL-L 0.2047 0.2677 0.2992 0.2835
   DL-M 0.2047 0.2677 0.2677 0.2835
   DL-H 0.2047 0.2205 0.2677 0.2835
favg (mm–1)
   FBP 0.3160 0.3188 0.3229 0.3213
   AV-30 0.2959 0.2989 0.3032 0.3017
   AV-50 0.2759 0.2788 0.2831 0.2825
   AV-80 0.2293 0.2314 0.2358 0.2355
   AV-100 0.1849 0.1895 0.1924 0.1934
   DL-L 0.2980 0.3019 0.3080 0.3066
   DL-M 0.2858 0.2926 0.2992 0.2987
   DL-H 0.2638 0.2764 0.2848 0.2845

NPS, noise power spectrum; FBP, filtered back projection; AV, adaptive statistical iterative reconstruction-Veo; DL-L, deep learning image reconstruction-low; DL-M, deep learning image reconstruction-medium; DL-H, deep learning image reconstruction-high.