Table 3.
IVUS Findings in Lesions Meeting vs Not Meeting OPTIVUS Criteria (per Lesion Basis)
OPTIVUS Criteria |
P Value | ||
---|---|---|---|
Meeting | Not Meeting | ||
Entire study population | (n = 1,246) | (n = 798) | |
Proximal reference vessel area, mm2 | 15.5 ± 5.3 (n = 1,041) | 16.9 ± 5.9 (n = 641) | <0.001 |
Proximal reference lumen area, mm2 | 7.9 ± 3.1 | 8.8 ± 3.5 | <0.001 |
Minimum stent area, mm2 | 5.7 ± 2.0 | 5.5 ± 2.1 | 0.03 |
Distal reference vessel area, mm2 | 9.3 ± 4.7 (n = 1,203) | 10.4 ± 5.6 (n = 769) | <0.001 |
Distal reference lumen area, mm2 | 5.2 ± 2.2 | 6.7 ± 2.9 | <0.001 |
Thrombus or protrusion | 158 (12.7) | 102 (12.8) | 0.95 |
Incomplete stent appositiona | 389 (31.2) | 354 (44.4) | <0.001 |
Dissection | 68 (5.5) | 28 (3.5) | 0.04 |
Stent length ≥28 mm | (n = 664) | (n = 564) | |
Proximal reference vessel area, mm2 | 16.9 ± 5.3 (n = 523) | 17.1 ± 5.6 (n = 443) | 0.59 |
Proximal reference lumen area, mm2 | 8.6 ± 3.4 | 8.7 ± 3.3 | 0.91 |
Minimum stent area, mm2 | 5.4 ± 1.8 | 5.5 ± 2.0 | 0.32 |
Distal reference vessel area, mm2 | 8.3 ± 4.1 (n = 645) | 10.0 ± 5.1 (n = 545) | <0.001 |
Distal reference lumen area, mm2 | 4.5 ± 1.7 | 6.6 ± 2.6 | <0.001 |
Thrombus or protrusion | 92 (13.9) | 83 (14.7) | 0.67 |
Incomplete stent appositiona | 251 (37.8) | 246 (43.6) | 0.04 |
Dissection | 39 (5.9) | 22 (3.9) | 0.11 |
Stent length <28 mm | (n = 582) | (n = 234) | |
Proximal reference vessel area, mm2 | 14.2 ± 5.0 (n = 518) | 16.5 ± 6.7 (n = 198) | <0.001 |
Proximal reference lumen area, mm2 | 7.1 ± 2.6 | 9.2 ± 3.9 | <0.001 |
Minimum stent area, mm2 | 6.1 ± 2.2 | 5.6 ± 2.3 | 0.003 |
Distal reference vessel area, mm2 | 10.5 ± 5.0 (n = 558) | 11.4 ± 6.6 (n = 224) | 0.03 |
Distal reference lumen area, mm2 | 5.9 ± 2.5 | 6.9 ± 3.5 | <0.001 |
Thrombus or protrusion | 66 (11.3) | 19 (8.1) | 0.17 |
Incomplete stent appositiona | 138 (23.7) | 108 (46.2) | <0.001 |
Dissection | 29 (5.0) | 6 (2.6) | 0.12 |
Values are mean ± SD. Categorical variables are presented as number and percentage.
Abbreviations as in Table 1.
Incomplete stent apposition was defined as the presence of blood flow between stent struts and vessel wall.