Knupfer 2001.
Study characteristics | |||
Patient Sampling | Cross‐sectional study, consecutive sampling | ||
Patient characteristics and setting |
Sample size: 135 Gestational age: preterm (23–36 weeks) Race/ethnicity: "Caucasians" (presumed white) and Asians Prior phototherapy: in some infants Setting: obstetrics department and NICU at the University of Leipzig, Leipzig, Germany |
||
Index tests |
TcB Device: BiliChek Site of TcB measurement: forehead |
||
Target condition and reference standard(s) |
Target condition: neonatal hyperbilirubinaemia TSB measurement method: diazo method (unclear if reference standard interpreted without knowledge of results of index test) |
||
Flow and timing |
Time interval between tests: not reported Number of paired results: 400 |
||
Comparative | |||
Notes | . | ||
Methodological quality | |||
Item | Authors' judgement | Risk of bias | Applicability concerns |
DOMAIN 1: Patient Selection | |||
Was a consecutive or random sample of patients enrolled? | Yes | ||
Was a case‐control design avoided? | Yes | ||
Did the study avoid inappropriate exclusions? | Yes | ||
Could the selection of patients have introduced bias? | Low risk | ||
Are there concerns that the included patients and setting do not match the review question? | Low concern | ||
DOMAIN 3: Reference Standard | |||
Is the reference standards likely to correctly classify the target condition? | Yes | ||
Were the reference standard results interpreted without knowledge of the results of the index tests? | Unclear | ||
Could the reference standard, its conduct, or its interpretation have introduced bias? | Low risk | ||
Are there concerns that the target condition as defined by the reference standard does not match the question? | Low concern | ||
DOMAIN 4: Flow and Timing | |||
Was there an appropriate interval between index test and reference standard? | Yes | ||
Did all patients receive the same reference standard? | Yes | ||
Were all patients included in the analysis? | Yes | ||
Could the patient flow have introduced bias? | Low risk |