Figure 5.

Figure 5.

Central mean arterial pressure (cMAP) assessments using the cardiac scale compared with a reference device, the SphygmoCor XCEL™ brachial blood pressure cuff. Left, Scatter plot with a linear regression line for the scale-derived and reference device cMAP measurements (n=252 data pairs from 88 participants; r=0.75; mean absolute percent error, 9.6%). Right, Bland-Altman analysis plotting the bias (mean error of cardiac scale minus reference device) of −0.14 mm Hg and 95% limits of agreement (LOA) of −26.1 to 25.8 mm Hg (dotted lines).