Skip to main content
. 2023 May 16;22:47. doi: 10.1186/s12938-023-01111-0

Table 6.

Default targets for the SOLVe algorithm as used during the in vivo experiments

Target Unit Target low Target high
SpO2 % 91 93
PETCO2 mmHg 30 60
Protective target region VT,pbw ml/kg 4 6
ΔP mbar 5 14
Protective acceptable region VT,pbw ml/kg 6 8
ΔP mbar 5 14
Protective compromise region VT,pbw ml/kg 4 6
ΔP mbar 14 20