TABLE III.

Normalized Error of Estimated Regional Uptake in the NEMA Phantom Using the LC-QSPECT Method With and Without Compensating for the Stray-Radiation-Related Noise

Normalized error (%)
Sphere diameter (mm) 10 13 17 22 28 37

LC-QSPECT method
(compensated for stray-radiation-related noise)
−62.02 −58.86 −29.84 −7.95 6.28 −10.03
LC-QSPECT method without
compensating for stray-radiation-related noise
61.43 47.97 55.31 37.63 41.20 34.31