State of the art in integrated biosensors for organ-on-a-chip applications

Yangzhi Zhu 1, Kalpana Mandal 1, Ana Lopez Hernandez 1, Satoru Kawakita 1, Wei Huang 1, Praveen Bandaru 1, Samad Ahadian 1, Han-Jun Kim 1, Vadim Jucaud 1, Mehmet R Dokmeci 1, Ali Khademhosseini 1
PMCID: PMC10193909  NIHMSID: NIHMS1893656  PMID: 37206309

Abstract

Organ-on-a-chip (OoC) models are bioengineered tissue constructs integrated with microfluidics that recapitulate the key features of the physiology of human organs and tissues with applications related to drug development and personalized medicine. The characterization of OoCs relies on conventional labor-intensive approaches despite the many years of research in the field. The physical environment of the tissue constructs, functionality, and metabolic activity of the cells must be monitored to ensure the behavior of the cells, and the cellular environments represent in vivo physiology. Current efforts focus on monitoring these parameters, particularly with in-line biosensors integrated with OoCs. In this review, we describe the recent advances in different biosensing modalities applied to monitor the environment and functionality of OoC models and offer suggestions for future directions in OoC applications.

Keywords: Organ-on-a-chip, Integrated biosensors, Biomarkers, Physical environment, Metabolic activity

Introduction

Organ-on-a-chip (OoC) models are bioengineered microfluidic devices fitted for the culture and analysis of cell cultures that recapitulate critical features of the physiology of human organs and tissues in vivo (i.e., architecture, cellular organization, and function) [2,3]. Since the development of a lung-on-a-chip published in 2010 [5], other OoC models have been established (gut, liver, kidney, heart, vessel, and blood–brain barrier [BBB]), whereas others have been recently reported (skin, bone, muscle, synovium, and eye) [3]. These OoCs have been extended to model disease states such as cardiac fibrosis [6], progeria [8], and different cancers [9]. In addition, there have been efforts to interconnect OoC units to recapitulate a human “body-on-a-chip” and complex organ–organ interactions [3]. However, most OoCs relyon microscopy and off-chip analytical techniques to assess and characterize the architecture, cellular organization, and function of the different organotypic models [11]. Because these techniques are destructive, in-line biosensors integrated with OoCs are promising alternatives to monitor and characterize OoCs in a minimally invasive manner, ultimately to make OoC models more functionally relevant to their in vivo counterparts.

The development of OoC models requires a microfluidic chip designed to recapitulate the specifications of an organ-specific microenvironment. The critical components that need to be considered include the spatial arrangements of cells, the presence of biomechanical forces, and tissue characterization techniques [12]. With the advent of stem cell technologies, personalized in vitro models that account for genotypic differences between individuals have also been developed [13]. These models may predict drug responses in a personalized manner, which holds tremendous potential for drug development and personalized medicine applications [13]. Nonetheless, it is crucial to monitor the physical parameters of the OoC’s microenvironment accurately because they directly affect cell viability, growth, organization, differentiation, and function. In addition, measuring biochemical markers indicating the metabolic and functional status of cultured cells is critical. Finally, in-line biosensors integrated with OoCs can be used to monitor the response to external stimuli (i.e. drug screening) and assess organotypic functionality under stringently controlled conditions.

Several biosensing methodologies have been integrated with OoC devices to monitor the microenvironment (pH, dissolved oxygen [DO], and temperature), cell metabolism and function (metabolic parameters, secreted biomarkers, organ activity, and barrier integrity), and response to external stimuli (electrical, mechanical, and drugs) (Figure 1). Integrating analytical biosensors to OoC platforms provides crucial information necessary to maintain a controllable and reproducible cell culture environment. Importantly, key features of these in-line biosensors integrated with OoCs are as follows: (1) compatibility with microfluidics, (2) high specificity and sensitivity, (3) label-free, and (4) minimally invasive [14]. Within this context, electrochemical (EC) and optical biosensing modalities are better suited to fulfill these requirements. Any fluctuation in physical parameters of the microenvironment might have detrimental effects on the cultured cells; therefore, continuous and real-time monitoring of the microenvironment is required to understand the physiological functionality of cultured cells within OoC models [15]. This review will describe different biosensing modalities used in OoC models, particularly related to monitoring physical parameters of the microenvironment and the physiological relevance of organotypic cell cultures using biochemical characterizations of cellular metabolism, function, and response to external stimuli.

Figure 1.

Figure 1

Integration of in-line sensors for monitoring the cellular microenvironment of OoC devices using (1) electrochemical biosensors (amperometry, enzymatic, impedance spectroscopy, and TEER), (2) optical biosensors (SPR and colorimetric detection), and (3) physical sensors (dissolved oxygen, pH, and temperature).

Types of biosensors for organ-on-a-chip applications

Electrochemical biosensors

EC biosensors have attractive features to monitor organotypic cell cultures within OoC devices because they can be easily integrated into OoC models and exhibit wide dynamic linear ranges and low limit of detection (LoD) [16]. EC biosensors using amperometry and voltammetry have been extensively used to monitor the physical parameters of the microenvironment and detect metabolic markers of OoC models. By detecting the current produced from the oxygen reduction reaction, EC sensors have been used to measure DO levels [17]. Continuous monitoring of DO concentrations was also realized with microelectrode arrays based on the amperometric concentration of oxygen reduction [18]. For enzymatic EC biosensors, the basic operation principle relies on a redox reaction between a bioreceptor (i.e. enzyme) immobilized onto the working electrode and the molecule to be detected. For example, Misun et al. [1] developed an amperometric sensor that relied on quantifying the increased number of free electrons produced by the redox reaction of glucose or lactate. This EC biosensor was integrated with a microfluidic hanging-drop device and provided continuous monitoring of glucose and lactate levels secreted from three-diemsional human colon cancer microtissues (Figure 2a). This enzyme-based EC biosensor featured high sensitivities of 322 ± 41 nA mM−1 mm−2 for glucose and 443 ± 37 nA mM−1 mm−2 for lactate.

Figure 2. Types of integrated biosensors for varied organ-on-a-chip applications.

Figure 2

(a). Electrochemical biosensing integrated in OoC applications. Left: design and assembly of the microfluidic hanging-drop network and sensor glass plug-ins; Middle: configuration of the biosensors and the corresponding mechanism illustration of the enzyme-based biosensors. Glucose oxidase (GOx) or lactate oxidase (LOx) transforms glucose or lactate to produce hydrogen peroxide, which is electrochemically detected on the electrode surface. Right: Glucose/lactate calibration curves showing the sensitivity and linearity [1]. (b). EIS-based biosensors used to monitor biomarkers of interest. Left: Schematic overview of the configuration and function of the muscle-on-a-chip. Middle: On-site multiplexed sensing system with sandwich immunoassay as the detection mechanism. Right: Calibration curves for IL-6 and TNF-α. Curves show limits of detection of 8 ng/mL and 2 ng/mL, respectively [4]. (c). Optical sensing system for real-time detecting of oxygen in OoC devices. Left: schematic of the sensing module. Right: Response of the sensor to DO in solutions [7]. (d). Multiplexed biophysical sensing platform in monitoring of temperature and pH for OoC applications. Left: Integrated automated multiorgan-on-a-chip and sensing platform. Middle: schematic diagram of biomimetic human heart-liver-on-chips with biosensing modules. Right: Continuous measurements of temperature and pH within the integrated heart-liver-on-chips [10]. Abbreviation: PDMS: polydimethylsiloxane; and iPSC: induced pluripotent stem cells.

EC impedance spectroscopy (EIS) is a powerful technique to detect biomolecules. EIS-based biosensors have been used in conjunction with antibodies or aptamers to detect specific biomarkers of interest. These biosensors exhibit high levels of selectivity because of the specific recognition of biomarkers (analyte) by antibodies or aptamers (bioreceptor). Briefly, self-assembling monolayers are functionalized onto gold electrodes. Next, specific primary antibodies or aptamers are immobilized on the electrodes via a carbodiimide cross-linking chemistry facilitated by (1-Ethyl-3-(3-dimethylaminopropyl)carbodiimide). Using EIS-based biosensors, Ortega et al. [4] detected interleukin (IL)-6 and tumor necrosis factor (TNF)-α using a labeled modality. After sample incubation, enzyme-linked secondary antibodies were used to detect the analytes of interest, where specific substrates reduced H2O2 to 2H2O and released a number of electrons directly proportional to the concentration of the analyte. An LoD of 8 ng/mL and 2 ng/mL was achieved for IL-6 and TNF-α, respectively (Figure 2b). Zhang et al. [10] used EIS sensors to monitor liver (albumin and glutathione S-transferase α (GST-α)) and heart (creatine kinase MB [CK-MB]) biomarkers in a multi-OoC platform (liver and heart). LoDs achieved were 0.09 (albumin), 0.01 (GST-α), and 0.024 (CK-MB) ng/mL.

Electrodes were used to characterize OoC models with a biological barrier, such as the BBB, air-liquid interface in lung-on-a-chip models, eye-on-a-chip barrier models, and the intestinal epithelial barrier of gut-on-a-chip models. The assessment of barrier function and integrity is critical when culturing endothelial or epithelial lining layers to establish the functionality of organotypic cell cultures. Transepithelial/transendothelial electrical resistance (TEER) is the measurement of electrical resistance across a cellular barrier. TEER is the most widely used parameter that characterizes barrier integrity because it is a label-free, rapid, and highly sensitive measurement [19,20]. Maoz et al. [20] demonstrated that TEER values could be used for real-time monitoring of biological function and response to therapeutic drugs. Park et al. integrated TEER-measuring electrodes into a BBB platform using 4 gold electrodes deposited on polycarbonate substrates to monitor dynamic changes in TEER of a BBB layer for at least one week of culture [21,22].

Optical biosensors

Optical biosensors can be used to detect biological parameters by studying different light properties (resonance, reflectance, polarization, and plasmon effect). They hold great promise for OoC applications because of their straightforward integration with microfluidics, label-free modality, and minimally invasive nature. Label-free optical sensors include surface plasmon resonance (SPR), total internal reflection-based sensors, photonic crystals, ring resonators, and waveguide fibers. Among different optical biosensors, SPR sensors are attractive because of their ability to measure molecules in real time [23]. In SPR biosensors, polarized light directed to the transducing surface provokes a plasmon effect and subsequently changes the refractive index affecting the resonance angle of the reflected light, which can be detected upon biomolecule binding [24]. Other technologies such as nanohole arrays (NHAs) have been incorporated into SPR sensors to enhance their sensitivity. NHAs have shown extraordinary optical transmission phenomenon because of the excitation of plasmons by coupling light at normal incidence. The light confinement in the nanometer scale allows the NHAs to be highly sensitive to surface refractive index changes induced by biomolecules’ binding to the biosensor surface. For OoC applicaitons, these changes can be readily monitored by tracking the variations of the transmission spectra, providing real-time sensitive detection and quantification of the biomolecules in a label-free manner. In this regard, SPR-based NHA was incorporated in a microfluidic cell module to detect vascular endothelial growth factor with an LoD of 145 pg/mL [25].

Optical approaches can also be used to monitor the physical environment of the cells in a minimally invasive manner. For example, Shaegh et al. [7] developed a microfluidic-based optical pH sensing platform based on detecting light absorbed by phenol red in the culture media. As the pH of the solution increased, the light absorbed through the culture medium also increased, which was used to monitor pH in the media. An oxygen sensor was also included by integrating a luminescence dye within the chip (Figure 2c) [7,23]. Optical measurement of DO was based on the photoluminescence quenching effect of oxygen, which decreased the intensity of luminophores and depended on the partial pressure of oxygen. Ruthenium- and metalloporphyrin-based fluorophores were primarily used because of their high quenching constants. Optical techniques based on fluorescence probes can be used for H2O2 monitoring. For example, carbon nanodots can be treated with hydroethidine, where the resulting material can react with O2 with high selectivity and generate products that exhibited an emission peak at 610 nm. This technology was able to achieve an LoD of 100 nM [26]. In another example, Fang et al. [27] developed a fluorescence sensing approach for H2O2 and hydroxyl radicals in living cells, with lysozyme acting as a fluorescence signal unit, for which they achieved an LoD of 0.2 mM.

Although all examples presented are a practical starting point for the field of in-line sensing of OoC systems, there is still a long way to achieve an optimal system adapted to the specific conditions in each case. This will involve optimizing the proposed biosensors herein described and their miniaturization to allow their integration in automated and controlled microfluidic platforms. In this regard and because of the materials, they are fabricated with electrical and optical biosensors, which are preferred for their susceptibility to being reduced in terms of size and weight, thanks to the nano and microfabrication techniques boosted during the last 30 years for telecommunication applications. Those fabrication advances are promising tools in integrated biosensors for in situ detection and point-of-care diagnoses.

Microfluidic-based biosensors are leading technologies that enhance biosensing performances by controlling different features such as the fluid transport rate, reagent mix, and sample volume, ultimately leading to an increased sensing sensitivity. In addition, microfluidic technology is the pillar of the OoC technology concept. It permits the development of organoids imitating mechanical, chemical, physical, and physiological features of living tissues. As microfluidic fabrication and experimentation methods are developed and incorporated into the OoC systems, this cutting-edge technology comes closer to reproduce tissue and organ characteristic with higher reliability.

Monitoring physical parameters of the microenvironment within OoCs

The microenvironment within OoC devices plays a crucial role in cellular maturation, viability, differentiation, and function. The continuous, real-time monitoring of the microenvironment allows the control of the performance and physiological functionality of cultured cells within OoC models [15]. A wide range of physical parameters of the microenvironment has been monitored in OoC devices, including pH, DO, temperature, carbon dioxide, nutrient content, osmolarity, humidity, surface tension, mechanical strain, and flow rates [17]. Any fluctuation in these parameters may affect the interpretation of studies conducted with OoC models and drug development (discovery, screening, and toxicity). For example, real-time measurement of DO is essential to monitor cellular function, as mitochondrial respiration could be lost when exposed to a hypoxic microenvironment. In addition, osmolality and pH affect cell viability and growth as they regulate the transport of water and nutrients through the cell membrane. Mattes et al. [28] developed an near-infrared reflextance spectroscopy-based biosensor to monitor the osmolality and pH in the culture medium of a bioreactor cultured with cells loaded on microcarriers. More recently, a multisensor-integrated multi-OoC system was developed by Zhang et al. [10], in which the team reported a fully integrated multiplexed biophysical sensing platform to monitor pH, oxygen, and temperature in a heart-liver-on-chip model for more than 5 days (Figure 2d).

Monitoring biochemical parameters of cellular metabolism, function, and response to stimuli

Biochemicals and proteins released from cells provide essential information about their viability, metabolic activity, and functionality while evaluating their biological responses to environmental conditions. Determining the metabolic activity mainly consists of monitoring glucose, lactate, or oxygen levels (Table 1). Recently, a liver-on-a-chip model was developed using a microwell bioreactor, and oxygen uptake was monitored to reflect mitochondrial dysfunction using two-frequency phase modulation of tissue-embedded phosphorescent microprobes [29]. Mitochondrial dysfunction leads to a decrease in oxygen uptake and an increase in lactate production, which was detected using an integrated automated microfluidic switchboard that captured the cellular shift from oxidative phosphorylation to glycolysis. In addition, Poudineh et al. [30] designed an aptamer-based biosensor to monitor glucose and insulin levels simultaneously and in real time. They achieved picomolar sensitivity using real-time enzyme-linked immunosorbent assay that used bead-based fluorescence, and an aptamer–quencher complex, in a microfluidic chip (Table 1). Microbeads were functionalized with fluorophore-conjugated glucose aptamers, and upon binding to glucose, the displacement of the competitor strand, which produced a fluorescent signal, was measured. The sandwich immune assay was used for insulin detection, where microbeads were functionalized with anti-insulin capture antibodies and phycoerythrin-conjugated detection.

Table 1.

Representative integrated biosensors in organ-on-a-chip devices.

Organ-on-a-chip Motivation Biosensing Materials Biomarkers Ref

Gut-on-a-chip Immunocompetent gut model recapitulates inflammation-induced leakiness Luminex-ELISA N/A IL-1β, Mip-1α, Mip-1 β, TNFα, IL-6, IL-8 [36]
Heart and breast-on-a-chip Dual organoid interaction and cardiac toxicity EC impedance spectroscopy Gold-based microelectrode array with aptamers immobilized on the surface Troponin T, CK-MB, HER-2 [31]
Blood–brain barrier Real-time measurement of TEER to characterize BBB-on-a-chip Impedance spectroscopy Polycarbonate substrate coated with 3 nm titanium and 25 nm gold TEER [22]
Blood–brain barrier Real-time measurement of TEER to characterize iPSC-based BBB-on-a-chip Impedance spectroscopy Multielectrode array (Pt) TEER electrodes (Au) TEER [19]
Muscle-on-a-chip Multiplexed in situ monitoring of 3D muscle microtissue on a chip for IL-6 and TNF-α Amperometric sensing system Functionalized screen-printed gold electrodes IL-6, TNF-α [4]
Liver-on-a-chip Cellular stress and disfunction monitoring Amperometric sensing system GOx- or LOx-coated electrodes Glucose, lactate [37]
Pancreatic microtissue Cellular stress and disfunction monitoring ELISA N/A Insulin, glucose [38]
Liver-on-a-chip and heart-on-a-chip Highly sensitive and long-term organoid monitoring TEER-MEA MEA connectors (Pt) TEER electrodes (Au) Albumin, glutathione S-transferase α, creatine kinase [10]
Colon-on-a-chip Real-time and in situ detection of the tissue secretion Enzyme-based biochemical sensor GOx- or LOx-coated electrodes Glucose, lactate [1]
Cardiac organoids Detection of cell–cell communications of 3D tissue organoids 3D-self-rolled biosensor array PEDOT:PSS modified microelectrodes over a polymer substrate Electrical signals (current/voltage potential) [39]
Neurons Measurement of electrical activity from neurons Passive electrode recording Complementary metal–oxide–emiconductor MEAs Electrical signals (current/voltage potential) [40]

ELISA, enzyme-linked immunosorbent assay; MEA, microelectrode array; 3D, three-dimensional; PEDOT: PSS, poly(3,4-ethylenedioxythiophene): poly (sodium 4-styrenesulfonate).

The cell secretome exhibits many unique biological signatures of specific cells. Therefore, quantitative detection of secreted biomarkers in complex biological media from cells holds promise in monitoring information regarding the status of the cells (Table 1). For example, cytokines such as ILs, interferons, and TNF-α are quantified to assess organoid functions, in particular for immune cell responses, cancer progression, and cell–cell signaling [31]. Ortega et al. [4] designed a muscle-on-a-chip model that was characterized by the released myokines, namely IL-6 and TNF-α, which play an essential role in muscle repair. Inline gold electrode biosensors detected these biomarkers with ng/ml sensitivity. Similarly, EIS biosensors were used to study the crucial role of breast and cardiac organoid interactions in response to chemotherapeutic drugs by measuring troponin T and CK-MB(cardiac biomarkers) and human epidermal growth factor receptor 2 (breast cancer biomarkers) with a sensitivity of 0.1 pg/mL [31]. Furthermore, matrix metalloproteinases (MMPs) are used as biomarkers for oxidative stresses and cardiovascular diseases and in basement membrane protein degradation that results in BBB leakage [32]. In that context, human fibro-blast–like synoviocytes from patients with rheumatoid arthritis were isolated and cultured in the microchambers, and IL-6, IL-8, MMP-1, and MMP-3 secretion was monitored using a noninvasive light scattering biosensing approach [33]. Riahi et al. [34] developed a bead-based microfluidic EC immunosensor for in-line detection of transferrin and albumin secreted from a liver-on-a-chip model. The platform allowed biomarker monitoring for 5 days. Finally, the cell secretome includes exosomes, which contain cell-specific signatures that participate in cell–cell communications. Zhou et al. [35] detected transmembrane proteins CD63 that were immobilized onto gold electrode surfaces from the exosome-mediated aptamer-based signaling, showing detection sensitivity using 100-fold fewer exosomes than a commercial assay, and incorporated into a microfluidic system.

Conclusions and future directions

Despite the advances in the fields of OoC systems and sensors, efforts to integrate sensors with OoCs are only beginning to emerge. The goal of the OoC systems is to create healthy or diseased tissue models that can represent in vivo tissue models. While building OoC systems, one needs to make sure that (i) the physical environment of the tissues is representative of the native conditions and (ii) cellular functionality and behavior are similar to their native counterparts. To address the former criterion, physical sensors are required, and for the second criterion, there is a need for biochemical sensors where both sensing platforms should be placed ideally in very close proximity to the OoC systems. Without such understanding, doubts exist while correlating the outcome of the measurements to external stimuli. The conventional endpoint monitoring methods are invasive and fail to provide sufficient data for understanding (1) the dynamics of proliferation, differentiation, response to stimuli of organoids, and tissue constructs grown within an OoC system; and (2) how to control and validate experimental outcomes using OoC systems. Although several sensing modalities have been integrated with OoCs to validate that the physical and biochemical microenvironment is akin to the in vivo physiological microenvironment—both in the structural organization of cell types and biochemical cues—the unmet need to monitor the status of OoCs during experimentation remains to be fully addressed. Notably, only physical sensors monitoring pH, oxygen, and temperature have been integrated with OoCs for real-time monitoring of the cellular microenvironment. In contrast, cell-secreted molecules, used to monitor cellular functionality, are monitored offline at single point measurements, which does not allow the dynamic characterization of cell behavior. Therefore, monitoring biomolecules in a real-time fashion remains warranted to validate that OoC models behave similarly to their native analogs. One issue with biosensors is that they tend to saturate rapidly, and the measurements are limited to a single use. As such, future work will need to focus on developing regeneration methods to multiuse these biosensors for continuous monitoring of specific soluble molecules and biomarkers. With the current trend in creating OoCs with multiple organ types, future efforts will need to be concentrated on the multiplexed detection of biomarkers from several OoCs. Other issues of biosensors such as biofouling, reliability, and scalability are challenges yet to be addressed specifically to OoC applications. Once these issues are addressed, only then the real-time monitoring of biomolecules will be realized. It is very desirable to have a single multisensor platform that can do both physical and biochemical sensing functions in an automated manner. Furthermore, knowing the various needs and properties of the OoC systems, we believe that each multisensory platform needs to be customized for the intended OoC application. Finally, we believe that the development of modular sensing systems which allow plug-and-play and allow replacement of sensors required because of biofouling and sensor failure would be of immense value. In addition, the modular approach would also benefit the OoC field to add additional OoC components to the systems when needed. The use of on-chip microfluidic valves will allow automation of the sensing processes and pave the ground for translating the OoC devices for high-throughput applications, such as drug development and screening.

Acknowledgements

The authors gratefully acknowledge funding by the National Institutes of Health, USA (CA214411, AR074234, GM126571, TR003148) and the Office of the Secretary of Defense, United States and were accomplished under agreement number W911NF-17–3-003. The views and conclusions contained in this document are those of the authors and should not be interpreted as representing the official policies, either expressed or implied, of the Office of the Secretary of Defense or the U.S. Government. The U.S. Government is authorized to reproduce and distribute reprints for government purposes notwithstanding any copyright notation herein.

Abbreviations

BBB

Blood–brain barrier

OoC

Organ-on-a-chip

TEER

Transepithelial/transendothelial electrical resistance

DO

Dissolved oxygen

EC

Electrochemical

LoD

Limit of detection

EIS

Electrochemical Impedance Spectroscopy

GST-α

Glutathione S-transferase α

SPR

Surface plasmon resonance

NHAs

Nanohole arrays

ELISA

Enzyme-linked immunosorbent assay

IL

Interleukins

CK-MB

Creatine kinase-MB

TNF

Tumor necrosis factor

MMP

Matrix metalloproteinases

Footnotes

Declaration of competing interest

The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.

References

Papers of particular interest, published within the period of review, have been highlighted as:

* of special interest

  • 1. Misun PM, Rothe J, Schmid YRF, Hierlemann A, Frey O: Multi-analyte biosensor interface for real-time monitoring of 3D microtissue spheroids in hanging-drop networks. Microsyst Nanoeng 2016, 2:16022. * This article comprehensively discusses the recent organ-on-a-chip technologies in recreating organ functions in-vitro, tissue function, and multi-organ interactions. It highlights the future challenges that need to be addressed for the clinical translation of OoC systems, including new materials, biosensing, scalability, and validation.
  • 2. Zhang B, Korolj A, Lai BFL, Radisic M: Advances in organ-on-a-chip engineering. Nature Rev Mater 2018, 3:257–278. * This work summarizes the new challenges, opportunities, and expertise of OoC technology in accelerating fundamental and translational applications and points out the possible directions that OoC systems could be pushed forward.
  • 3.Low LA, Mummery C, Berridge BR, Austin CP, Tagle DA: Organs-on-chips: into the next decade. Nat Rev Drug Discov 2020, 20:345–361. [DOI] [PubMed] [Google Scholar]
  • 4.Ortega MA, Fernandez-Garibay X, Castano AG, De Chiara F, Hernandez-Albors A, Balaguer-Trias J, Ramon-Azcon J: Muscle-on-a-chip with an on-site multiplexed biosensing system for in situ monitoring of secreted IL-6 and TNF-alpha. Lab Chip 2019, 19:2568–2580. [DOI] [PubMed] [Google Scholar]
  • 5.Huh D, Matthews BD, Mammoto A, Montoya-Zavala M, Hsin HY, Ingber DE: Reconstituting organ-level lung functions on a chip. Science 2010, 328:1662–1668. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Kong M, Lee J, Yazdi IK, Miri AK, Lin YD, Seo J, Zhang YS, Khademhosseini A, Shin SR: Cardiac fibrotic remodeling on a chip with dynamic mechanical stimulation. Adv Healthc Mater 2019, 8, e1801146. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Mousavi Shaegh SA, De Ferrari F, Zhang YS, Nabavinia M, Binth Mohammad N, Ryan J, Pourmand A, Laukaitis E, Banan Sadeghian R, Nadhman A, et al. : A microfluidic optical platform for real-time monitoring of pH and oxygen in microfluidic bioreactors and organ-on-chip devices. Biomicrofluidics 2016, 10. 044111–044111. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Ribas J, Zhang YS, Pitrez PR, Leijten J, Miscuglio M, Rouwkema J, Dokmeci MR, Nissan X, Ferreira L, Khademhosseini A: Biomechanical strain exacerbates inflammation on a progeria-on-a-chip model. Small 2017, 13. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Sontheimer-Phelps A, Hassell BA, Ingber DE: Modelling cancer in microfluidic human organs-on-chips. Nat Rev Cancer 2019, 19:65–81. [DOI] [PubMed] [Google Scholar]
  • 10.Zhang YS, Aleman J, Shin SR, Kilic T, Kim D, Mousavi Shaegh SA, Massa S, Riahi R, Chae S, Hu N, et al. : Multisensor-integrated organs-on-chips platform for automated and continual in situ monitoring of organoid behaviors. Proc Natl Acad Sci U S A 2017, 114:2293–e2302. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Rothbauer M, Ertl P: Emerging biosensor trends in organ-on-a-chip. Springer; Berlin Heidelberg; 2020:1–12. [DOI] [PubMed] [Google Scholar]
  • 12.Wu Q, Liu J, Wang X, Feng L, Wu J, Zhu X, Wen W, Gong X: Organ-on-a-chip: recent breakthroughs and future prospects. Biomed Eng Online 2020, 19:9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Ebert AD, Liang P, Wu JC: Induced pluripotent stem cells as a disease modeling and drug screening platform. J Cardiovasc Pharmacol 2012, 60:408–416. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Luka G, Ahmadi A, Najjaran H, Alocilja E, DeRosa M, Wolthers K, Malki A, Aziz H, Althani A, Hoorfar M: Microfluidics integrated biosensors: a leading technology towards lab-on-a-chip and sensing applications. Sensors 2015, 15:30011–30031 (Basel, Switzerland). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Clarke GA, Hartse BX, Niaraki Asli AE, Taghavimehr M, Hashemi N, Abbasi Shirsavar M, Montazami R, Alimoradi N, Nasirian V, Ouedraogo LJ, et al. : Advancement of sensor integrated organ-on-chip devices. Sensors 2021, 21 (Basel). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Karimi-Maleh H, Karimi F, Alizadeh M, Sanati AL: Electrochemical sensors, a bright future in the fabrication of portable kits in analytical systems. Chem Rec 2020, 20:682–692. [DOI] [PubMed] [Google Scholar]
  • 17.Kilic T, Navaee F, Stradolini F, Renaud P, Carrara S: Organs-on-chip monitoring: sensors and other strategies. Microphysiological Systems 2018, 2. [Google Scholar]
  • 18. Ferrari E, Palma C, Vesentini S, Occhetta P, Rasponi M: Integrating biosensors in organs-on-chip devices: a perspective on current strategies to monitor microphysiological systems. Biosensors 2020, 10 (Basel). * This work describes the integration of OoC technology and human iPSC-derived cells that enabled the creation of a BBB-on-a-chip that recapitulates the complex human BBB and may serve as a platform for in vitro disease modeling, drug screening, and personalized medicine. An EI approach was taken to measure TEER to characterize the barrier function via integrated platinum multi-electrode arrays and gold electrodes.
  • 19.Vatine GD, Barrile R, Workman MJ, Sances S, Barriga BK, Rahnama M, Barthakur S, Kasendra M, Lucchesi C, Kerns J, et al. : Human iPSC-derived blood-brain barrier chips enable disease modeling and personalized medicine applications. Cell Stem Cell 2019, 24:995–1005. e1006. [DOI] [PubMed] [Google Scholar]
  • 20.Maoz BM, Herland A, Henry OYF, Leineweber WD, Yadid M, Doyle J, Mannix R, Kujala VJ, FitzGerald EA, Parker KK, et al. : Organs-on-Chips with combined multi-electrode array and transepithelial electrical resistance measurement capabilities. Lab Chip 2017, 17:2294–2302. [DOI] [PubMed] [Google Scholar]
  • 21.Henry OYF, Villenave R, Cronce MJ, Leineweber WD, Benz MA, Ingber DE: Organs-on-chips with integrated electrodes for trans-epithelial electrical resistance (TEER) measurements of human epithelial barrier function. Lab Chip 2017, 17:2264–2271. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Park TE, Mustafaoglu N, Herland A, Hasselkus R, Mannix R, FitzGerald EA, Prantil-Baun R, Watters A, Henry O, Benz M, et al. : Hypoxia-enhanced Blood-Brain Barrier Chip recapitulates human barrier function and shuttling of drugs and antibodies. Nat Commun 2019, 10:2621. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Zhou J, Qi Q, Wang C, Qian Y, Liu G, Wang Y, Fu L: Surface plasmon resonance (SPR) biosensors for food allergen detection in food matrices. Biosens Bioelectron 2019, 142:111449. [DOI] [PubMed] [Google Scholar]
  • 24.Mejía-Salazar JR, Oliveira ON: Plasmonic biosensing. Chem Rev 2018, 118:10617–10625. [DOI] [PubMed] [Google Scholar]
  • 25.Noori A, Ashrafi SJ, Vaez-Ghaemi R, Hatamian-Zaremi A, Webster TJ: A review of fibrin and fibrin composites for bone tissue engineering. Int J Nanomed 2017, 12:4937. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Gao X, Ding C, Zhu A, Tian Y: Carbon-dot-based ratiometric fluorescent probe for imaging and biosensing of superoxide anion in live cells. Anal Chem 2014, 86:7071–7078. [DOI] [PubMed] [Google Scholar]
  • 27.Liu F, Bing T, Shangguan D, Zhao M, Shao N: Ratiometric fluorescent biosensing of hydrogen peroxide and hydroxyl radical in living cells with lysozyme–silver nanoclusters: lysozyme as stabilizing ligand and fluorescence signal unit. Anal Chem 2016, 88:10631–10638. [DOI] [PubMed] [Google Scholar]
  • 28.Mattes RA, Root D, Sugui MA, Chen F, Shi X, Liu J, Gilbert P: Real-time bioreactor monitoring of osmolality and pH using near-infrared spectroscopy. 2009.
  • 29.Han J, Kim B, Shin J-Y, Ryu S, Noh M, Woo J, Park J-S, Lee Y, Lee N, Hyeon T: Iron oxide nanoparticle-mediated development of cellular gap junction crosstalk to improve mesenchymal stem cells’ therapeutic efficacy for myocardial infarction. ACS Nano 2015, 9:2805–2819. [DOI] [PubMed] [Google Scholar]
  • 30.Poudineh M, Maikawa CL, Ma EY, Pan J, Mamerow D, Hang Y, Baker SW, Beirami A, Yoshikawa A, Eisenstein M, et al. : A fluorescence sandwich immunoassay for the real-time continuous detection of glucose and insulin in live animals. Nat Biomed Eng 2020. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31. Lee J, Mehrotra S, Zare-Eelanjegh E, Rodrigues RO, Akbarinejad A, Ge D, Amato L, Kiaee K, Fang Y, Rosenkranz A, et al. : A heart-breast cancer-on-a-chip platform for disease modeling and monitoring of cardiotoxicity induced by cancer chemotherapy. Small 2020, e2004258. * This article summarizes a dual-organ-on-a-chip model, consisting of two-chamber units housing iPSC-derived cardiac cells and breast cancer cells with in-line, multiplexed. EC immuno-aptasensors to monitor cell-secreted biomarkers. The platform provides an efficient means to assess chemotherapy-induced cardiotoxicity.
  • 32.Liu W, Song J, Du X, Zhou Y, Li Y, Li R, Lyu L, He Y, Hao J, Ben J, et al. : AKR1B10 (Aldo-keto reductase family 1 B10) promotes brain metastasis of lung cancer cells in a multi-organ microfluidic chip model. Acta Biomater 2019, 91:195–208. [DOI] [PubMed] [Google Scholar]
  • 33.Rothbauer M, Höll G, Eilenberger C, Kratz SRA, Farooq B, Schuller P, Olmos Calvo I, Byrne RA, Meyer B, Niederreiter B, et al. : Monitoring tissue-level remodelling during inflammatory arthritis using a three-dimensional synovium-on-a-chip with non-invasive light scattering biosensing. Lab Chip 2020, 20:1461–1471. [DOI] [PubMed] [Google Scholar]
  • 34.Riahi R, Shaegh SAM, Ghaderi M, Zhang YS, Shin SR, Aleman J, Massa S, Kim D, Dokmeci MR, Khademhosseini A: Automated microfluidic platform of bead-based electrochemical immunosensor integrated with bioreactor for continual monitoring of cell secreted biomarkers. Sci Rep 2016, 6. 24598–24598. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Zhou Q, Rahimian A, Son K, Shin DS, Patel T, Revzin A: Development of an aptasensor for electrochemical detection of exosomes. Methods 2016, 97:88–93. [DOI] [PubMed] [Google Scholar]
  • 36.Gjorevski N, Avignon B, Gerard R, Cabon L, Roth AB, Bscheider M, Moisan A: Neutrophilic infiltration in organ-on-a-chip model of tissue inflammation. Lab Chip 2020, 20:3365–3374. [DOI] [PubMed] [Google Scholar]
  • 37.Bavli D, Prill S, Ezra E, Levy G, Cohen M, Vinken M, Vanfleteren J, Jaeger M, Nahmias Y: Real-time monitoring of metabolic function in liver-on-chip microdevices tracks the dynamics of mitochondrial dysfunction. Proc Natl Acad Sci Unit States Am 2016, 113:E2231–E2240. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38.Misun PM, Yesildag B, Forschler F, Neelakandhan A, Rousset N, Biernath A, Hierlemann A, Frey O: In vitro platform for studying human insulin release dynamics of single pancreatic islet microtissues at high resolution. Advanced Biosystems 2020, 4:1900291. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 39.Kalmykov A, Huang C, Bliley J, Shiwarski D, Tashman J, Abdullah A, Rastogi SK, Shukla S, Mataev E, Feinberg AW, et al. : Organ-on-e-chip: three-dimensional self-rolled biosensor array for electrical interrogations of human electrogenic spheroids. Science Advances 2019, 5, eaax0729. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40.Muller J, Ballini M, Livi P, Chen Y, Radivojevic M, Shadmani A, Viswam V, Jones IL, Fiscella M, Diggelmann R, et al. : High-resolution CMOS MEA platform to study neurons at subcellular, cellular, and network levels. Lab Chip 2015, 15:2767–2780. [DOI] [PMC free article] [PubMed] [Google Scholar]

RESOURCES