Table 1.

Rules-based filter for invalid physiological data.

Rules Filter for invalid data Range
1 To prevent “floor” artifacts (eg, electrode loses contact with skin) and “ceiling” artifacts (circuit is overloaded)—EDAa not in a valid range 0.05 to 60 µSb
2 EDA changes too quickly—EDA slope not in a valid range −10 to +10 µS/second
3 Skin temperature suggests the EDA sensor is not being worn—skin temperature not in a valid range 30 to 40 °C
4c HRd not in a valid range 25 to 250 bpme
5 Transitional data surrounding segments identified as invalid via the preceding rules—account for transition effects Within 5 seconds

aEDA: electrodermal activity.

bµS: microsiemens.

cAddition to the algorithm used by Kleckner et al [52].

dHR: heart rate.

ebpm: beats per minute.