Table 1.
Methods implemented for acquisitions of experimental data used to set the model basic physical aspects and parameters in the 21 selected articles.
| First author name | Year | Experimental conditions | ||
|---|---|---|---|---|
| Treatment Machine - Manufacturer: [beam energies] | Phantom | Detectors | ||
| Francois | 1988 | Conventional x-rays – Philips [200lVp], Theratron 80 – AECL [60Co], Sagittaire – CGR [25 MV] | Water Box/Slabs | 0.6 cc IC |
| McParland | 1992 | Clinac 6/100 - Varian [6 MV] | Water Box/Slabs | 0.6 cc IC |
| van der Giessen | 1993 | Theratron-780 – Theratronics [60Co] | Water Box/Slabs | 0,6 cc graphite IC |
| van der Giessen | 1994 | Saturne 41 – GE/CGR [6 MV, 10 MV], Saturne 25 – CGR [10 MV, 23MV], Therac 6 – Theratronics [6 MV] | Water Box/Slabs | 0.6 cc nylon IC |
| Diallo | 1996 | Theratron 780-AECL [60Co], Orion-GE [4MV] and others not specified | In-air | Film |
| van der Giessen | 2001 | See van der Giessen 1993 and van der Giessen 1994 | N/A | N/A |
| Benadjaoud | 2012 | Alcyon II – CGR [60Co], Clinac 2300 C/D – Varian [6MV, 20MV], Novalis Tx – Varian [6 MV] | Water Box/Slabs | TLD-700 |
| Taddei | 2013 | Clinac 2100 – Varian [6 MV], Artiste – Siemens [6 MV] | Anthropomorphic | TLD-100 |
| Veres | 2014 | See (6) | N/A | N/A |
| Bezin | 2015 | Novalis Tx - Varian [6 MV] | In-air | TLD-700 |
| Jagetic | 2015 | SL25 - Elekta [6 MV], Synergy-S/N:151892 - Elekta | Water Box/Slabs | Diamond detector, IC |
| Sánchez-Nieto | 2015 | Primus - Siemens [15 MV, 18 MV], Mevatron - Siemens [23 MV], Clinac - Varian [15 MV], Synergy - Elekta [6MV, 15MV] | N/A | TLD-700, TLD-100 |
| Hauri | 2016 | Clinac iX - Varian [6 MV], TrueBeam - Varian [6 MV] | Water Box/Slabs | IC |
| De Saint-Hubert | 2017 | TrueBeam - Varian [6 MV], Clinac 2100C/D - Varian [6 MV], Hi-Art - Accuray | Anthropomorphic | MCP-N TLD |
| Schneider | 2017 | Saturne 43 - GE [6 MV, 12MV, 20 MV], SL25 - Elekta [6 MV, 18 MV, 25 MV], Clinac 21 iX - Varian [6MV], Synergy - Elekta [6MV], Oncor Avant-Garde-Siemens [6MV] | Water Box/Slabs | RPL, OSL, diamond detector, TLD |
| Yoon | 2017 | Versa HD - Elekta [6MV FFF, 6MV,10MV FFF, 10 MV], TomoTherapy – Accuray | Anthropomorphic | TLD-100 |
| Gallagher | 2018 | See (53) | N/A | N/A |
| Hauri | 2019 | TrueBeam - Varian [6MV,15 MV] | Anthropomorphic, see (54) | IC and TLD100/100H/700/700H |
| Schneider | 2019 | PTS – Elekta [6 MV] | Water Box/Slabs | IC, diode dosimeter |
| Wilson | 2020 | See (55) | N/A | N/A |
| Sánchez-Nieto | 2022 | N/A | Anthropomorphic | N/A |
IC, ionization chamber; TLD, thermoluminescent dosimeter; RPL, radiophotoluminescent dosimeter; OSL, optically stimulated luminescence.