Table 4.
Accuracy of two-dimensional TEE and three-dimensional TEE diagnosis of thrombus and pannus on left mechanical valve obstruction
| Diagnosis methods | True positive (n) | True negative (n) | False positive (n) | False negative (n) | Sensitivity (%) | Specificity (%) | Accuracy (%) | PPV (%) | NPV (%) |
|---|---|---|---|---|---|---|---|---|---|
| Diagnosis of thrombus | |||||||||
| 2-dimensional TEE | 19 | 2 | 1 | 26 | 42.2 | 66.7 | 43.8 | 95.0 | 7.1 |
| 3-dimensional TEE | 33 | 8 | 3 | 4 | 89.2 | 72.7 | 85.4 | 91.7 | 66.7 |
| Diagnosis of pannus | |||||||||
| 2-dimensional TEE | 2 | 19 | 2 | 25 | 7.4 | 90.5 | 43.8 | 50 | 43.2 |
| 3-dimensional TEE | 8 | 33 | 0 | 7 | 53.3 | 100 | 85.4 | 100 | 82.5 |
NPV, negative predictive value; PPV, positive predictive value; TEE, transesophageal echocardiography.