Table 2. Instrumentation and operational settings.

ICP-MS
Instrument Magnetic sector-based ICP-MS (AttoM, Nu Instruments, UK)
ICP forward power 1300 W
Argon gas flow rate
Coolant gas 13.0 L min−1
Auxiliary gas 1.0 L min−1
Monitored isotopes 107Ag and 197Au
Dwell time 50 μs
Detector ETP full size multiplier
Detection system Pulse counting mode with attenuator system
Desolvation sample introduction system
Instrument Aridus II (Teledyne Cetac Technologies, USA)
Sample uptake rate 0.2 mL min−1
Spray chamber temperature 110°C
Membrane oven temperature 160°C
Sweep gas Ar (4.8 L min−1)
Laser ablation system
Instrument Nd:YAG laser (MOPA 266–200 mW, CryLas, Germany)
Wavelength 266 nm (FHG)
Pulse width 1 ns
Fluence 0.2 to 3.0 J cm−2
Repetition rate 10 Hz
Carrier gas He (0.60 L min−1)
Makeup gas Ar (1.0 L min−1) mixed after the ablation cell using T-piece connector
Scanning speed 100 μm s−1
Laser beam size 5 μm