Table 2.

Adjusted correlations for the maximum push-out force with bulk morphological descriptors of the initial bone–implant domain. Significant correlations are marked with a star.

Material Time N BV/TV BIC ABIC Vscrew,ext BA/BVb κnorm,pos
Titanium any 17 0.06 −0.30 0.19 −0.53* −0.12 0.09
PEEK any 11 0.84* 0.48 0.43 0.23 −0.82* −0.88*
Mg-10Gd any 10 0.43 0.78* 0.74* −0.72* −0.77* −0.54
any 4 weeks 14a 0.50 0.51 0.39 0.47 −0.64* −0.66*
any 8 weeks 11 0.06 0.27 0.26 −0.45 −0.55 −0.25
any 12 weeks 17 0.17 0.20 0.05 0.20 −0.11 −0.01
Titanium 4 weeks 6 −0.37 0.01 0.47 −0.90* 0.63 0.37
Titanium 8 weeks 5 −0.86 −0.96* 0.24 −0.97* 0.17 0.42
Titanium 12 weeks 6 0.03 −0.25 −0.18 −0.07 0.81 0.70
PEEK 12 weeks 5 0.84 0.14 0.05 0.16 −0.35 −1.00*
Mg-10Gd 12 weeks 5 0.30 0.90* 0.77 −0.76 −0.39 −0.26
any any 42 0.50* 0.32 0.29 −0.06 −0.65* −0.53*

* significant at p < 0.05.

a

N = 13 for the calculation of the Vscrew, exterior correlation.

b

Interfaces to the screw and to narrow vessels were excluded from the evaluation of the surface area.