Skip to main content
. 2023 May 22;28(10):4225. doi: 10.3390/molecules28104225

Table 1.

Summary of selected liquid chromatography and mass spectrometry conditions.

Mobile phase. Binary system of
0.1% Formic acid in H2O (A) and ACN (B)
ESI source Positive ESI
High purity N2 gas Drying gas at 10 L/min with pressure (60 psi)
0.4 mL/min.
Elution time: 45 min.
Agilent Zorbax eclipse plus C18 Column Length 150 mm ESI temperature: 350 °C
ID 4.6 mm Capillary voltage: 4000 V
Particle size 3.5 μm Modes Mass scan, MS2 and MS3
Temperature 22 ± 1 °C Collision gas High purity N2
Gradient elution system Time in min. % ACN Analytes FBB and its metabolites
0 5 Mass parameters Fragmentor voltage (FV): 145 V
Amplitude: 1.25 V
20 60
25 65
30 65
45 5