Skip to main content
. Author manuscript; available in PMC: 2023 May 28.
Published in final edited form as: Phys Med Biol. 2022 Mar 11;67(6):10.1088/1361-6560/ac508e. doi: 10.1088/1361-6560/ac508e

Figure 3.

Figure 3.

Distortion measured in the phantom using the B0 mapping technique at BW=210 Hz (blue circles), the same curve scaled by 0.42 (grey squares), and the distortion measured manually at BW=500 Hz (orange triangles). This data shows that increasing bandwidth will accurately reduce distortion by a predictable amount.